[Faulty internal tube in a co-axial ventilation tube system: cause of a massive postoperative hypercapnia].
- J-H Günther, P Börning, and L Bahlmann.
- Institut für Anästhesie und Intensivmedizin, Katholische Hospitalvereinigung Weser-Egge gGmbH, Brenkhäuserstr. 71, 37671, Höxter, Deutschland. j.guenther@khwe.de
- Anaesthesist. 2013 Mar 1;62(3):197-200.
AbstractThis article presents the case of a patient with massive postoperative hypercapnia during mechanical ventilation in the intensive care unit (ICU). With normal tidal volumes and clearly visible chest movements, adequate findings with regard to auscultation, oxygenation and correct respirator settings, no cause for the increasing hypercapnia was initially found; however, replacement of the respirator led to a return to normal carbon dioxide levels. When checking the replaced respirator a service technician found the cause of the respirator failure: the internal tube of the co-axial ventilation system was faulty leading to an increased dead space and rebreathing of carbon dioxide.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.