J Clin Monit Comput · Feb 2014
Bench test assessment of mainstream capnography during high frequency oscillatory ventilation.
- Caroline M Hartdorff, Marc van Heerde, and Dick G Markhorst.
- Pediatric Intensive Care Unit, VU University Medical Center, PO BOX 7057, 1007 MB, Amsterdam, The Netherlands.
- J Clin Monit Comput. 2014 Feb 1; 28 (1): 63-6.
AbstractTo assess the feasibility, stability and predictability of pCO2 measurement (PETCO2) using a main stream capnograph in a high frequency oscillatory ventilation circuit. A commercially available capnograph was mounted into a high frequency oscillatory ventilator patient circuit, adjustable CO2 flow was introduced into an artificial lung and the output of the CO2 sensor assessed under varying ventilator settings. Influence of oxygen content, pressures, heat and moisture were recorded. A linear relationship between CO2 flow rate and PETCO2 was found. Varying ventilator settings influenced the measurements, but the results for PETCO2 remained within a range of 1.5 mmHg above or under then mean measurement value. Measurements remained stable despite humidification, heat, pressure amplitudes or mean airway pressure changes. From this bench test, we conclude it is feasible to measure PETCO2 using a main stream capnograph during high frequency oscillatory conditions, these measurements were stable during the experiment. Changes in CO2 production or output can be detected. The system may prove to be of clinical value, but further in vivo measurements are warranted.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.