• Eur Spine J · Jul 2021

    Multicenter Study

    Metallosis is commonly associated with magnetically controlled growing rods; results from an independent multicentre explant database.

    • RushtonPaul R PPRP0000-0002-3387-9416Great North Children's Hospital, Royal Victoria Infirmary, Newcastle upon Tyne, UK. prp.rushton@gmail.com., Simon L Smith, David Fender, Andrew J Bowey, Michael J Gibson, and Thomas J Joyce.
    • Great North Children's Hospital, Royal Victoria Infirmary, Newcastle upon Tyne, UK. prp.rushton@gmail.com.
    • Eur Spine J. 2021 Jul 1; 30 (7): 1905-1911.

    PurposeDetermine the incidence of metallosis around MAGEC rods.MethodsA multicentre explant database was searched to identify cases with complete intraoperative findings at rod removal. Surgeons removing rods detailed the presence or absence of tissue metallosis associated with rods. More recently surgeons measured the 'length' of tissue metallosis. Prior to rod disassembly, the majority underwent testing with an external remote controller (ERC). The impact of clinical and explant variables on metallosis was assessed.ResultsSixty-six cases were identified. Mean age at insertion was 8.1 ± 2.3 years with mean duration of implantation 37.6 ± 15.1 months. Tissue metallosis was noted at revision surgery in 52/66 cases (79%). Metallosis was noted more commonly when rods were removed during fusion surgery than rod removal/exchange (97% vs. 58% (p =  < 0.01)). The mass at insertion was greater in cases with metallosis (25.9 ± 7.8 kg vs. 21.1 ± 6.2 kg, p = 0.04). Length of tissue metallosis was reported for 45 rods, median 9 cm (range 1-25). Metallosis was noted in 43/59 (73%) rods that produced no force and 22/30 (73%) rods that produced some force on ERC activation (p = 0.96). Wear debris was found within the actuator in all rods, and all but 3 rods had damaged O-rings.ConclusionMAGEC rods are associated with tissue metallosis in the majority of cases. It is seen with functional rods as well as failed rods and appears related to wear debris within the actuator and high rates of O-ring failure. Until the implications of metal debris in children are known, we urge caution with the use of this implant.© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH, DE part of Springer Nature.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…