J. Cardiothorac. Vasc. Anesth. · Jul 2021
Continuous Noninvasive Arterial Pressure Monitoring for Transcatheter Aortic Valve Replacement.
- Shu Y Lu and Adam A Dalia.
- Department of Anesthesiology, Pain Medicine, and Critical Care Medicine, Massachusetts General Hospital, Harvard Medical School, Boston, MA; Department of Critical Care Medicine, Massachusetts General Hospital Heart Center, Massachusetts General Hospital, Harvard Medical School, Boston, MA.
- J. Cardiothorac. Vasc. Anesth. 2021 Jul 1; 35 (7): 2026-2033.
ObjectivesThe objective of the present study, which was conducted in patients undergoing transcatheter aortic valve replacement, was to compare continuous noninvasive arterial pressure measured with the ClearSight device (Edwards Lifesciences, Irvine, CA) with invasive radial artery pressure used as the reference method. The authors hypothesized that the ClearSight device is an accurate, precise, safe, and efficient method for arterial blood pressure measurement comparable with an invasive radial arterial line.DesignThe study included the retrospective review of 20 consecutive patients scheduled for elective transcatheter aortic valve replacement with the SAPIEN 3 transcatheter heart valve (Edwards Lifesciences) at a single tertiary academic hospital, who underwent monitoring with both the ClearSight device and an invasive radial arterial pressure line. The patients underwent transcatheter aortic valve replacement from October to December 2019.SettingSingle tertiary academic medical center.ParticipantsThe study comprised 20 patients, with 2,243 unique blood pressure data points.InterventionsNone.Measurements And Main ResultsA statistically significant correlation between the invasive radial arterial pressure line and the ClearSight device was observed for systolic blood pressure (correlation coefficient 0.86; p < 0.001), diastolic blood pressure (correlation coefficient 0.56; p < 0.001), and mean arterial pressure (correlation coefficient 0.78; p < 0.001). Bland-Altman analysis was used to assess the agreement of systolic blood pressure, diastolic blood pressure, and mean arterial pressure between the two methods. Results for systolic blood pressure between the arterial line and ClearSight device were as follows: bias = 9.8 ± 10.1, percentage bias = 7.6%, and mean error = 15.8%. Results for diastolic blood pressure between the arterial line and ClearSight device were as follows: bias = -5.9 ± 7.8, percentage bias = 10.7%, and mean error = 28.4%. Results for mean arterial pressure between the arterial line and ClearSight device were as follows: bias = 0.3 ± 7.4, percentage bias = 0.4%, and mean error = 18.3%. The concordance rates of systolic blood pressure, diastolic blood pressure, and mean arterial pressure were 100%, 95.1%, and 98.8%, respectively.ConclusionsThe accuracy, agreement, and precision of the ClearSight device were convincing for mean arterial pressure, systolic blood pressure, and diastolic blood pressure for patients with severe aortic stenosis undergoing elective transcatheter aortic valve replacement.Copyright © 2021 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.