    Growth guidance system for early-onset scoliosis: comparison of experimental and retrieval wear.

    • Vaneet Singh, Joshua Simpson, Jeremy Rawlinson, and Nadim Hallab.
    • Medtronic Inc., Memphis, TN, USA. vaneet.singh@medtronic.com
    • Spine. 2013 Aug 15;38(18):1546-53.

    Study DesignLaboratory study conducted using an in vitro wear simulator with a growth guidance system. Analysis of variance performed to compare in vitro specimens (n = 6) with in vivo retrieval components (n = 5).ObjectiveTo characterize the stainless steel, wear debris potential of a spinal growth guidance system by developing an in vitro model and validating tested implants with retrospectively obtained retrievals.Summary Of Background DataGrowth enabling, surgical treatments have been developed to provide fusionless options for patients with early-onset scoliosis. There exist few data regarding the wear debris associated with such spinal systems.MethodsIn this study, we determined in vitro wear from the stainless steel components of the SHILLA™ Growth Guidance System. An analogue lumbar spine model was adapted from ISO 12189:2008 to assess the growth guidance system. In a multistation wear simulator, 6 assembled constructs were tested under displacement control for 5 million cycles (Mc) with diluted bovine serum, and the wear was measured gravimetrically at end of the test. The components were compared quantitatively for wear scar depth with retrieved growth guidance implants (n = 5), and qualitatively for wear, corrosion, and other surface damage.ResultsThe average total wear rate over 5 Mc was 0.39 ± 0.13 mm/Mc (3.12 ± 1.01 mg/Mc) with an average particle size of 1.3 μm in equivalent circular diameter. Prominent wear scars were noticed on both the tested and retrieved specimens with no statistical difference in the wear scar depths of the tested and retrieved components when set and multiaxial screws when compared collectively.ConclusionAn in vitro wear analysis for a spinal growth guidance system was conducted using a novel protocol and validated against retrieved implants. This is the first study establishing a baseline value for the wear of "growth enabling" devices for the treatment of early-onset scoliosis.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…