Biomechanical evaluation of S2 alar-iliac screws: effect of length and quad-cortical purchase as compared with iliac fixation.
- Joseph R OʼBrien, Warren Yu, Brian E Kaufman, Brandon Bucklen, Kanaan Salloum, Saif Khalil, and Manasa Gudipally.
- *George Washington University MFA, Washington, DC; and †Globus Medical, Inc., Valley Forge Business Center, Audubon, PA.
- Spine. 2013 Sep 15;38(20):E1250-5.
Study DesignA biomechanical study conducted on cadaveric specimens.Objective(1) To compare the biomechanical strength of the S2 alar-iliac (S2AI) screw to traditional iliac fixation and (2) to examine the effect of length and trajectory on the S2AI screw.Summary Of Background DataA recent technique to attain spinal fixation distal to S1 pedicle screws is the S2AI screw using either an open or a percutaneous approach with an altered S2 alar screw trajectory to obtain purchase in the ilium. A novel modification of the S2AI screw is placement with bicortical purchase in the ilium (quad-cortical screw). This may allow for a shorter-length screw with equivalent biomechanics.MethodsSeven human cadaveric spines (L2-Pelvis) were fixed at L2 proximally and the pubis distally. Pedicle screws were placed from L3-S1 with S2AI screw lengths of 65-mm, 80-mm, or 90-mm iliac screws. S2AI screws were tested with and without quad-cortical purchase. Each specimen was tested on the 6 degrees of freedom spine simulator. A load control protocol with an unconstrained pure moment of 10 Nm was used in flexion-extension, lateral bending, and axial rotation for a total of 3 load/unload cycles. The range of motion was normalized to the intact cadaveric spine (100%).ResultsAll the instrumented constructs significantly reduced range of motion compared with the intact spine. The L3-S1 construct was statistically significantly less stable than all instrumented constructs in flexion-extension. There was statistically no significant difference between the S2AI screws of all lengths and the iliac screw constructs with offset connectors.ConclusionS2AI screws are biomechanically as stable as the test constructs using iliac screws in all loading modes. Sixty-five-millimeter S2AI screws were biomechanically equivalent to 90-mm iliac screws and 80-mm S2AI screws. Quad-cortical purchase did not statistically significantly improve the biomechanical strength of S2AI screws.Level Of EvidenceN/A.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.