• Ann. Clin. Biochem. · May 2017

    Stability of ionized calcium measurements at concentrations less than 0.3 mmol/L by point-of-care blood gas instruments: application for postfilter calcium quality control in patients with citrate anticoagulation during extracorporeal circulation.

    • Maria Averina and Raymond Jakobsen.
    • 1 Department of Laboratory Medicine, University Hospital of North Norway, Tromsø, Norway.
    • Ann. Clin. Biochem. 2017 May 1; 54 (3): 398-401.

    AbstractBackground Measurements of ionized calcium (Ca2+) at concentrations less than 0.3 mmol/L are required for postfilter control in patients who receive extracorporeal circulation with sodium citrate anticoagulation. This study evaluates the stability of the Ca2+ measurements at such concentrations. Methods The stability of the Ca2+ measurements was tested by measuring daily the external standard Qualicheck concentration 3 s7950, Radiometer (0.22-0.25 mmol/L) by blood gas instruments ABL800 and ABL90, Radiometer. Two different Ca2+ membrane lots were tested for the usual membrane lifetime of 12 weeks at ABL800 instruments. For the ABL90 instrument, the sensor cassette (with Ca2+ membrane and electrode) was replaced after four weeks as required. Results We observed over 40% Ca2+ increase within the usual 12 weeks lifetime of the Ca2+ membrane at the ABL800 instruments. Measurements of Ca2+ at concentrations less than 0.3 mmol/L were within acceptable limits for both ABL800 and ABL90 instruments when Ca2+ membrane and sensor cassette were replaced after four weeks. Conclusions For ABL800 instruments, it is necessary to use an extra quality control (<0.3 mmol/L) in addition to the usual quality controls to monitor Ca2+ measurements below 0.3 mmol/L. The acceptable stability of the Ca2+ measurements can be achieved by the Ca2+ membrane and sensor cassette replacement after four weeks. If the usual 12 weeks of Ca2+ membrane lifetime is maintained, it may result in a clinically significant overestimation of Ca2+ by ABL800 instruments.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…