-
J. Heart Lung Transplant. · Jul 2016
Evaluation of von Willebrand factor with a fully magnetically levitated centrifugal continuous-flow left ventricular assist device in advanced heart failure.
- Ivan Netuka, Tomáš Kvasnička, Jan Kvasnička, Ingrid Hrachovinová, Peter Ivák, František Mareček, Jana Bílková, Ivana Malíková, Michaela Jančová, Jiří Malý, Poornima Sood, Kartik S Sundareswaran, Jean M Connors, and Mandeep R Mehra.
- Department of Cardiac Surgery, Institute for Clinical and Experimental Medicine, Prague, Czech Republic; Second Department of Surgery, First Faculty of Medicine, Charles University of Prague, Prague, Czech Republic. Electronic address: ivan.netuka@ikem.cz.
- J. Heart Lung Transplant. 2016 Jul 1; 35 (7): 860-7.
BackgroundContemporary continuous-flow left ventricular assist devices (CF-LVADs) are associated with degradation of von Willebrand factor (vWF) high-molecular-weight multimers (HMWMs), a critical factor supporting platelet function. We hypothesized that the HeartMate 3 fully magnetically levitated LVAD, designed to reduce circulatory shear stress, favorably influences these hemostatic parameters.MethodsFifteen consecutive HeartMate 3 LVAD patients were compared with 11 consecutive HeartMate II controls. Serial plasma samples were collected pre-implant and on Days 2, 7, 30 and 45 post-operatively. Changes in vWF HMWMs were evaluated by 2 independent, study-blind hematologists and confirmed using densitometry-based computerized software. Ristocetin cofactor (RiCO) and vWF antigen (vWF Ag) were measured using standard protocols with enzyme-linked immunosorbent assay.ResultsHeartMate 3 patients and HeartMate II controls had a mean age of 67.3 ± 1.4 and 52.8 ± 2.5 years, respectively (INTERMACS Profiles 2 to 4 in 93.3% and 91%, respectively). HeartMate 3 group demonstrated a significantly greater preservation of HMWMs compared with the HeartMate II group, with the most prominent decrease occurring by Day 2 post-operatively and sustained through 45 days (71.94% vs 31.16%, p = 0.001). Laboratory values (normalized to baseline) for RiCO activity, vWF Ag and RiCO:vWF Ag ratio remained in the functional range with no statistically significant differences observed between groups.ConclusionThe HeartMate 3 LVAD is associated with enhanced hemocompatibility compared with the HeartMate II LVAD, as demonstrated by the improved preservation of vWF HMWMs, In contrast, effects on HMWM degradation appeared to be dissociated from functional attributes. Further confirmation of these findings in randomized clinical trials is warranted.Copyright © 2016 International Society for Heart and Lung Transplantation. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.