Catheter Cardiovasc Interv · Sep 2016
Comparative Study Observational StudyClinical outcome and paravalvular leakage of the new balloon-expandable Edwards Sapien 3 valve in comparison to its predecessor model (Edwards Sapien XT) in patients undergoing transfemoral aortic valve replacement.
- Wolfgang Bocksch, Benita Grossmann, Tobias Geisler, Martin Steeg, Michal Droppa, Rezo Jorbenadze, Michael Haap, Meinrad Gawaz, and Suzanne Fateh-Moghadam.
- Department of Medicine III (Cardiovascular Disease), Eberhard-Karls-Universitaet Tuebingen, Germany. Wolfgang.bocksch@sana.de.
- Catheter Cardiovasc Interv. 2016 Sep 1; 88 (3): 466-75.
ObjectivesThe aim of this study was to compare the 30-day procedural, clinical and echocardiographic outcome of the new balloon-expandable Edwards Sapien 3 (ES3) valve with the Edwards Sapien XT (ESXT).BackgroundPost-implant paravalvular leaks (PVL) after transfemoral aortic valve replacement (TAVR) resulting in residual aortic regurgitation (AR) are a major limitation for long term outcome. New TAVR-devices have to eliminate this problem.MethodsTransfemoral TAVR was performed in 209 consecutive intermediate-high-risk surgical patients (pts) with symptomatic aortic stenosis (ESXT n = 102, ES3 n = 107). Transthoracic echocardiography (TTE) and 3-dimensional computed tomography were used for valve size selection. Primary endpoint of the study was none/trace AR derived by TTE 30-days after TAVR.ResultsAll pts underwent successfully TAVR with a combined device success of 100/102 (99%) in ESXT and 107/107 (100%) in ES3 pts. Fluoroscopy time (ESXT 11.8 ± 0.5 min vs. ES3 10.0 ± 0.5 min, P = 0.003) and contrast (ESXT 188.9 ± 5.6 mL vs. ES3 170.4 ± 4.7 mL, P = 0.04) were significantly lower in ES3 patients. 30-day clinical events did not differ. Transvalvular mean pressure gradients were significantly reduced to 7.4 ± 0.8 mmHg after ESXT and to 10.1± 0.4 mmHg after ES3 implantation. After 30 days none/trace AR was found in 34.3% (n = 35) of all ESXT pts in contrast to 89.7% (n = 96) of all ES3 patients. Moderate-to-severe AR was found rarely (ESXT 2.9% vs. ES3 0%, P = 0.073).ConclusionsAlthough there was no significant difference in 30 day mortality, the newer ES3 valve reduced significantly residual paravalvular leakage. © 2016 Wiley Periodicals, Inc.© 2016 Wiley Periodicals, Inc.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.