    Polymer-on-metal or metal-on-polymer total disc arthroplasty: does it make a difference?

    • Parshia M Moghadas, Duncan E T Shepherd, David W L Hukins, and Aziza Mahomed.
    • School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham, United Kingdom. MoghadaP@adf.bham.ac.uk
    • Spine. 2012 Oct 1;37(21):1834-8.

    Study DesignMechanical testing of total disc arthroplasty (TDA).ObjectiveTo compare the friction between a polymer socket-on-metal ball and metal socket-on-polymer ball TDA.Summary Of Background DataA degenerate intervertebral disc can be replaced by TDA. The most common designs have a ball and socket articulation; the contact between the surfaces leads to friction. Friction needs to be minimized to prevent loosening and wear. One of the common material combinations in disc arthroplasty devices is the articulation of a metal socket on polymer ball. However, the combination of a polymer socket on metal ball (which is used in hip arthroplasty) has not been investigated for TDA.MethodsTDA models with either a polymer socket/metal ball or a metal socket/polymer ball were manufactured with ball radii of 10 and 14 mm, each with a radial clearance of 0.35 mm. Samples were tested using a spine simulator with a lubricant of diluted newborn calf serum. Each sample was subjected to an axial load of 1200 N; motions of flexion-extension, lateral bending, and axial rotation were then applied at frequencies of 0.25 to 2 Hz. Frictional torque was measured to compare the performance of the TDAs.ResultsThe frictional torque was found to be significantly higher for a disc with a metal socket/polymer ball than for a disc with a polymer socket/metal ball for both 10 and 14 mm radii in axial rotation, lateral bend, and extension. The frictional torque in flexion (0°-6°) was not found to be significantly different between the 2 different material combinations. However, when the flexion motion was reduced to 0° to 2°, frictional torque in the metal socket/polymer ball was found to be significantly higher than the polymer socket/metal ball.ConclusionTDA with a combination of a polymer socket/metal ball has lower friction than the conventional TDA with metal socket/polymer ball. This conclusion has implications in the design of TDA.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.