• Spine · Sep 2012

    In vivo functional performance of failed Prodisc-L devices: retrieval analysis of lumbar total disc replacements.

    • Darren R Lebl, Frank P Cammisa, Federico P Girardi, Timothy Wright, and Celeste Abjornson.
    • Hospital for Special Surgery, The Spine Care Institute, New York, NY 10021, USA. drlebl@alumni.stanford.edu
    • Spine. 2012 Sep 1;37(19):E1209-17.

    Study DesignA retrieval analysis of wear modes and fixation of lumbar total disc replacements (TDRs). Explanted Prodisc-L TDRs were prospectively collected during a 7-year period (2005-2011) and analyzed.ObjectiveTo assess the in vivo modes of wear and fixation of lumbar TDR with the Prodisc-L device.Summary Of Background DataInferior clinical outcomes and failure of lumbar TDR may occur because of suboptimal component fixation, wear properties, and impingement in a subset of patients. Posterior component TDR impingement has been demonstrated radiographically; however, despite its widespread use, the in vivo mechanical performance and fixation of the Prodisc-L device remain unknown.MethodsExplanted polyethylene and metallic (CoCrMo) components of Prodisc-L devices were examined by light stereo-microscopy (6X-31X), scanning electron microscopy, and energy-dispersive x-ray analysis from an international retrieval registry, with 13 participating surgeons.ResultsNineteen ProDisc-L devices from 18 patients (age, 44.7 ± 2.9 yr) following an index TDR at L4-L5 (n = 6), L5-S1 (n = 11), and unknown level (n = 2) were explanted for pain (n = 8), prosthesis subluxation/migration (n = 4), end plate collapse/subsidence (n = 3), polyethylene dislodgement (n = 3), and unknown (n = 2) after a mean length of implantation of 13.0 ± 3.9 months. Surface area of bony ongrowth was 9.6 ± 2.9% (range, 0%-52.5%). TDR burnishing was observed posteriorly consistent with component impingement in extension in 53% (8/15) (P < 0.02), more commonly than anterior 20% (3/15) lateral 20% (n = 3) (3/15) patterns. Circumferential burnishing was not observed. Posterior impingement was associated with 6° lordotic implants (P < 0.05) and 10-mm polyethylene size (P < 0.05). Backside wear occurred in 75% (9 of 12) of the disassembled implants and third-body wear was observed in 33% (5 of 15).ConclusionMetallic end plate burnishing was evident in a large percentage of clinically failed Prodisc-L TDR devices, most commonly posteriorly, consistent with impingement in extension. Long-term follow-up studies will evaluate the effects of the observed backside wear, third-body wear, and end plate impingement on clinical outcomes.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…