- Darren R Lebl, Frank P Cammisa, Federico P Girardi, Timothy Wright, and Celeste Abjornson.
- Hospital for Special Surgery, The Spine Care Institute, Spine and Scoliosis Surgery, New York, NY 10021, USA. drlebl@stanfordalumni.org
- Spine. 2012 Dec 15;37(26):2151-60.
Study DesignProspective retrieval analysis of Prodisc-C cervical total disc replacements (CTDRs) from 24 explanting surgeons during a 6-year period.ObjectiveTo determine the in vivo mechanical performance and fixation to bone of explanted Prodisc-C CTDRs.Summary Of Background DataThe nature and quantity of damage sustained by an implanted device has proven to be important in the prediction of clinical longevity. We hypothesized that retrieval analysis of the Prodisc-C will display characteristic modes of wear consistent with increased posterior angulation and translation of the functional spinal unit after resection of the discoligamentous anatomy.MethodsThirty CTDRs from 29 patients (mean age, 45.1 ± 1.9; range, 31-57 yr) after a mean length of implantation of 1.0 ± 0.2 years (range, 2 d-3.5 yr) were studied. Operative level was C4-C5 in 20% (6 of 30), C5-C6 in 47% (14 of 30), C6-C7 in 20% (6 of 30), and unknown in 13% (4 of 30). Polyethylene and metallic (cobalt chrome molybdenum [CoCrMo]) components were examined using light stereo-microscopy (6X-31X), scanning electron microscopy, and energy dispersive x-ray analysis.ResultsCTDRs were explanted for indications of axial pain (n = 9), radicular symptoms (n = 6), atraumatic loosening (n = 6), trauma (n = 5), metal allergy (n = 1), myelopathy (n = 1), hypermobility (n = 1), and unknown (n = 1). Surface area of ongrowth (mean = 7.2 ± 1.4%) was not associated with operative level (P = 0.37), surgeon-reported axial pain (P = 0.56), or atraumatic loosening (P = 0.93). Burnishing consistent with metallic endplate impingement was present in 80% (24 of 30) of retrieved CTDRs, most commonly in the posterior quadrant (P < 0.001). There was no association between implant height (P = 0.19) or depth (P = 0.17) and posterior impingement. Backside wear was not observed on any of the disassembled implants (0 of 16). Third-body wear occurred in 23% (7 of 30) and the donor site was confirmed by scanning electron microscope/energy dispersive x-ray analysis to be the porous-coated surface of the CTDR.ConclusionEarly clinical failures of Prodisc-C CTDRs display surface damage evidence of metal endplate-endplate impingement, most commonly posteriorly. Backside wear was not evident; however, third-body wear was found. Future studies will determine the clinical impact of these predominant modes of wear on long-term metal-on-polyethylene semiconstrained CTDR performance.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.