Case Reports
Unexpected Collection of Pulmonary Venous Blood from a Pulmonary Artery Catheter: A Case Report.
- Naomi Ono, Junko Nakahira, Sayuri Matsunami, Toshiyuki Sawai, and Toshiaki Minami.
- Department of Anesthesiology, Osaka Medical College, Takatsuki, Japan.
- Anesth Pain Med. 2016 Dec 1; 6 (6): e42621.
BackgroundPressure and waveform at the catheter tip are continuously monitored during catheterization of pulmonary artery to ensure accurate catheter placement. We present a case in which pulmonary venous blood was unexpectedly collected from the pulmonary artery catheter despite pulmonary artery pressure and waveform detection at the catheter tip, and describe the measures taken to correct the catheter placement.Case PresentationA 74-year-old male underwent mitral valve plasty for cardiac failure caused by mitral valve regurgitation. Preoperative transthoracic echocardiography showed no septal shunt. The pulmonary artery was catheterized through a sheath introducer in the right jugular vein, and the balloon was inflated after insertion of a 15-cm catheter. The catheter was advanced until a pulmonary artery waveform was detected and the pulmonary artery wedge pressure was 21 mmHg at end-expiration. The balloon was deflated and the catheter tip was pulled back 3 cm. Pulmonary artery waveforms and appropriate a and v waves were detected, and transesophageal echocardiography confirmed the location of the catheter tip in the right pulmonary artery. The first collected blood sample had an oxygen partial pressure of 358.8 mmHg, carbon dioxide partial pressure of 20.1 mmHg, and oxygen saturation of 99%, indicating pulmonary venous blood. The pulmonary artery catheter was pulled back 5 cm, but a second blood sample showed the same results. The catheter was pulled back a further 6 cm while the location of the catheter tip was monitored on X-ray fluoroscopy. Blood gas testing through the catheter tip showed oxygen saturation of 84.4 % and oxygen partial pressure of 41.6 mmHg. Surgery was performed uneventfully. Postoperative chest radiographs showed proper placement of the pulmonary artery catheter, but radiographs on postoperative day 1 showed over-insertion, although the insertion length was unchanged. The catheter was removed. The patient was discharged 2 months postoperatively.ConclusionsOur case highlights the fact that the tip of the pulmonary artery catheter can easily advance into a peripheral branch of the pulmonary artery and cause pulmonary venous blood to be sampled instead of pulmonary arterial blood. A variety of monitoring techniques are needed to confirm accurate catheter placement.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.