• J Neurointerv Surg · Apr 2020

    Diffusion weighted image documented cerebral ischemia in the postprocedural period following pipeline embolization device with shield technology treatment of unruptured intracranial aneurysms: a prospective, single center study.

    • Stylianos Pikis, Georgios Mantziaris, Vasileios Mamalis, Konstantinos Barkas, Antonios Tsanis, Stavroula Lyra, Kuriakos Karkoulias, Tigran Petrosyan, and Eftychios Archontakis.
    • Neurosurgery, 'Korgialenio Benakio' The Red Cross Hospital of Athens, Athens, Greece.
    • J Neurointerv Surg. 2020 Apr 1; 12 (4): 407-411.

    ObjectiveAcute ischemic stroke and silent cerebral infarctions following pipeline embolization device (PED) treatment of intracranial aneurysms have been estimated to occur in 3-6% and in 50.9-90% of patients respectively. The PED with Shield technology (PED-Shield) incorporates a surface phosphorylcholine polymer to reduce the thrombogenicity of the implant. We sought to determine the incidence of diffusion weighted image (DWI) documented cerebral ischemia after PED-Shield treatment of unruptured intracranial aneurysms.MethodsThis prospective study involved a single center series of consecutive patients treated for an unruptured intracranial aneurysm with the PED-Shield. All participants underwent clinical evaluation on admission, after the procedure, at discharge, and 30 days following treatment. Brain MRI was obtained within 72 hours of the procedure. Ischemic lesions identified on DWI sequences where examined as to their number, size, and location in relation to the procedure.ResultsOver 12 months, 33 patients harboring 38 intracranial aneurysms were treated with the PED-Shield in 36 procedures. Neither mortality nor clinically evident ischemic events were noted in the 30 day postprocedural period. DWI documented, silent cerebral ischemia occurred in six patients (18.18%) after six procedures (16.66%). No statistically significant risk factors for postprocedural silent cerebral ischemia were identified.ConclusionWe demonstrated a reduced rate of silent cerebral infarcts following PED-Shield treatment of intracranial aneurysms than previously reported with other endovascular treatment modalities and with the previous device generations. Further research is necessary to evaluate our results and to identify methods to reduce the incidence of postprocedural cerebral ischemic events.© Author(s) (or their employer(s)) 2020. No commercial re-use. See rights and permissions. Published by BMJ.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…