    Choice of faecal immunochemical test matters: comparison of OC-Sensor and HM-JACKarc, in the assessment of patients at high risk of colorectal cancer.

    • Caroline J Chapman, Ayan Banerjea, David J Humes, Jaren Allen, Simon Oliver, Abby Ford, Katie Hardy, Natasha Djedovic, Richard F Logan, and Joanne R Morling.
    • Eastern Hub, Bowel Cancer Screening Programme, A Floor West Block, QMC Campus, Nottingham University Hospitals NHS Trust, Nottingham, UK.
    • Clin Chem Lab Med. 2021 Mar 26; 59 (4): 721-728.

    ObjectivesCurrently, NICE recommends the use of faecal immunochemical test (FIT) at faecal haemoglobin concentrations (f-Hb) of 10 μg Hb/g faeces to stratify for colorectal cancer (CRC) risk in symptomatic populations. This f-Hb cut-off is advised across all analysers, despite the fact that a direct comparison of analyser performance, in a clinical setting, has not been performed.MethodsTwo specimen collection devices (OC-Sensor, OC-S; HM-JACKarc, HM-J) were sent to 914 consecutive individuals referred for follow up due to their increased risk of CRC. Agreement of f-Hb around cut-offs of 4, 10 and 150 µg Hb/g faeces and CRC detection rates were assessed. Two OC-S devices were sent to a further 114 individuals, for within test comparisons.ResultsA total of 732 (80.1%) individuals correctly completed and returned two different FIT devices, with 38 (5.2%) CRCs detected. Median f-Hb for individuals diagnosed with and without CRC were 258.5 and 1.8 µg Hb/g faeces for OC-S and 318.1 and 1.0 µg Hb/g faeces for HM-J respectively. Correlation of f-Hb results between OC-S/HM-J over the full range was rho=0.74, p<0.001. Using a f-Hb of 4 µg Hb/g faeces for both tests found an agreement of 88.1%, at 10 µg Hb/g faeces 91.7% and at 150 µg Hb/g faeces 96.3%. A total of 114 individuals completed and returned two OC-S devices; correlation across the full range was rho=0.98, p<0.001.ConclusionsWe found large variations in f-Hb when different FIT devices were used, but a smaller variation when the same FIT device was used. Our data suggest that analyser-specific f-Hb cut-offs are applied with regard to clinical decision making, especially at lower f-Hb.© 2020 Walter de Gruyter GmbH, Berlin/Boston.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.