• Anesthesia and analgesia · Oct 2021

    Anatomical In Vitro Investigations of the Pediatric Larynx: A Call for Manufacturer Redesign of Tracheal Tube Cuff Location and Perhaps a Call to Reconsider the Use of Uncuffed Tracheal Tubes.

    • Malaak Isa, Josef Holzki, Anna Hagemeier, Markus A Rothschild, and Charles J Coté.
    • From the Institute of Legal Medicine, Medical Faculty, University of Cologne, Germany.
    • Anesth. Analg. 2021 Oct 1; 133 (4): 894902894-902.

    BackgroundSome in vivo studies question the traditional "funnel-shaped" infant larynx; further anatomic examinations were warranted. Examination of fixative free fresh autopsy laryngeal and upper tracheal specimens and multiple measurements was needed to determine consistency between current tracheal tube designs and anatomic observations.MethodsLarynges from 19 males and 11 females (Caucasian term newborn to 126 months) were examined by the same forensic pathologist. Measurements included anterior/posterior (A/P) and transverse (T) diameters of the cricoid outlet (CO), interarytenoid diameter (IAD), cricothyroid membrane (CTM), distance from the vocal cords (VC) to CO (VC-CO), and calibration of the larynx lumen with uncuffed tracheal tubes as measuring rods. Assessment of "safe tracheal tube placement" was assessed using manufacturer recommended cuffed Microcuff (Kimberly-Clark, Koblenz, Germany) tubes.ResultsIn 77% (95% confidence interval [CI], 58-90) of specimens, the proximal end of the cuff was within the CO and in 23% even with or close to the CO. The VC-CO varied from 9.1 to 13.17 mm in infants, 11.55 to 15.17 mm in toddlers, and 13.19 to 18.34 mm in children. The A-P/T ratio of the CO was nearly 0.99 in most larynges; the IAD was greater than CO in all specimens. The CTM could be minimally distended in all specimens.ConclusionsFirst, despite being marketed as a safer tracheal tube design, the proximal end of the Microcuff cuff rested within or close to the cricoid cartilage theoretically increasing potential cuff-induced injury when using the VC markings for positioning. Our data suggest that the optimal cuff free distance (VC-CO) would be ~13.5 mm for a Microcuff internal diameter (ID) size 3.0, ~15 mm for size 3.5, and ~16 to 19 mm for greater sizes.Second, the CO was virtually circular in all specimens, suggesting that appropriately sized uncuffed tubes should provide an adequate seal in most neonates and toddlers, thus avoiding the potential for cuff-related necrosis injury.Third, the IAD was always greater than CO confirming that the narrowest point of the infant larynx is the nondistensible cricoid cartilage and not the easily distended glottis.Fourth, appropriately sized Microcuff tubes with the cuff deflated completely filled the lumen of the CO and proximal trachea in all specimens. Our data suggest the need for all manufacturers to further evaluate tracheal tube cuff locations and lengths in relation to the VC safe insertion markings, particularly for neonates and toddlers.Fifth, the CTM is minimally distensible, thus having important implications for emergency surgical airway access with most currently available emergency airway devices.Copyright © 2021 International Anesthesia Research Society.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…