• Critical care medicine · Nov 2002

    Comparative Study

    Pulse oximetry during low perfusion caused by emerging pneumonia and sepsis in rabbits.

    • Helmut D Hummler, Frank Pohlandt, and Axel R Franz.
    • Department of Pediatrics, Children's Hospital, Unversity of Ulm, Germany. helmut.hummler@medizin.uni-ulm.de
    • Crit. Care Med. 2002 Nov 1; 30 (11): 2501-8.

    ObjectiveThis study tested the effects of low perfusion caused by emerging sepsis on the reliability of a new pulse oximetry technology (Masimo SET; IVY 405T) compared with a standard pulse oximeter (Nellcor N-200).DesignRandomized trial.SettingUniversity animal research facility.SubjectsTwenty-six anesthetized, ventilated (Fio, 1.0), adult rabbits.InterventionsPneumonia/sepsis was induced by tracheal instillation of Escherichia coli. Oxygen saturation was measured by pulse oximetry (Spo ) and recorded continuously until death. Arterial oxygen saturation (Sao2) was measured hourly by oximetry and whenever Spo dropped to /=5% between devices occurred. Spo2 sensors were positioned at both forelegs and switched hourly.Measurements And Main ResultsThe total time of signal loss was longer with the N-200 vs. the IVY: 65 (4-299) mins vs. 7 (0-97) mins [median (range)], p < 0.001. Signal loss was more prevalent during the first 80% of the experimental time with the N-200 compared with the IVY. Nineteen of 26 animals had a total of 62 episodes of a falsely low Spo2 value with either one of the two devices associated with hemodynamic deterioration. Median bias (Spo2 - Sao2) was small, but variability of bias values increased toward the end of the experimental time with both devices.ConclusionsThe pulse oximeter equipped with Masimo SET was less prone to signal loss than the standard pulse oximeter in this sepsis model. Episodes of falsely low Spo2 readings may occur, and deviation of Spo2 from Sao2 may be increased with deteriorating hemodynamics with both devices.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,704,841 articles already indexed!

We guarantee your privacy. Your email address will not be shared.