J Spinal Disord Tech · Aug 2007
Comparative StudyComparative biomechanical analysis of an improved novel pedicle screw with sheath and bone cement.
- Tomoyuki Takigawa, Masato Tanaka, Hitoshi Konishi, Hisanori Ikuma, Haruo Misawa, Yoshihisa Sugimoto, Kazuo Nakanishi, Koichi Kuramoto, Keiichiro Nishida, and Toshifumi Ozaki.
- Department of Orthopaedic Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, Japan. takigawa2004@yahoo.co.jp
- J Spinal Disord Tech. 2007 Aug 1; 20 (6): 462-7.
Study DesignA human cadaveric biomechanical study of fixation strength of an improved novel pedicle screw (NPS) with cement and a conventional screw.ObjectiveTo clarify whether the NPS has adequate fixation strength without leakage in vertebrae with low bone quality.Summary Of Background DataThe fixation strength of pedicle screws decreases in frail spines of elderly osteoporotic patients. Augmentation of screw fixation with bone cement must be balanced against increased difficulty of screw removal and risk of cement leakage. We developed the NPS consisting of an internal screw and an outer sheath to mitigate the disadvantages of cement augmentation.MethodsThe T12 and L1 vertebrae obtained from 18 formalin preserved cadavers (11 males and 7 females; mean age, 82.7 y) were used. The mean bone mineral density was 0.39 +/- 0.14 g/cm2. The NPS was inserted into one pedicle of each vertebra and the control screw, a Compact CD2 screw, was inserted into the contralateral pedicle. Both screws were 6mm in diameter and 40 mm in length. Pull-out tests were performed at a crosshead speed of 10 mm/min. Cyclic loading tests were performed with a maximum 250 N load at 2 Hz until 30,000 cycles.ResultsCement leakage did not occur in any of the specimens tested. The mean maximum force at pull-out was 760 +/- 344 N for the NPS and 346 +/- 172N for the control screw (P < 0.01). Loosening of 50% of the screws was observed after 17,000 cycles of the NPS and after 30 cycles of the control screw. The hazard ratio of loosening was 19.6 (95% confidence interval 19.3-19.9) (P < 0.001).ConclusionsThe NPS showed a significantly higher mechanical strength than the control screw in both pull-out tests and cyclic loading tests. The NPS showed more than adequate strength without cement leakage.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.