• Anesthesia and analgesia · Jan 2022

    Comparative Study

    The Evaluation of a Noninvasive Respiratory Volume Monitor in Mechanically Ventilated Neonates and Infants.

    • Douglas B Atkinson, Brooke A Sens, Rachel S Bernier, Andrea D Gomez-Morad, Jasmin Imsirovic, and Viviane G Nasr.
    • From the Department of Anesthesiology, Critical Care and Pain Medicine, Boston Children's Hospital, Harvard Medical School, Boston, Massachusetts.
    • Anesth. Analg. 2022 Jan 1; 134 (1): 141148141-148.

    BackgroundThe respiratory volume monitor (RVM) (ExSpiron, Respiratory Motion Inc, Watertown, MA) uses thoracic impedance technology to noninvasively and continuously measure tidal volume (TV), respiratory rate (RR), and minute ventilation (MV). We aimed to validate the accuracy of the RVM to assess ventilation in neonates and infants by comparing it to spirometry.MethodsWe used the RVM and Respironics NM3 spirometer (Respironics NM3 Respiratory Profile Monitor, Philips Healthcare, Amsterdam, the Netherlands) to record simultaneous and continuous measurements of MV, TV, and RR. The RVM measurements, with and without external calibration, were compared to the Respironics NM3 spirometer using Bland-Altman analysis. The relative errors (Bland-Altman) between RVM and Respironics NM3 were calculated and used to compute individual patient bias, precision, and accuracy as the mean error, the standard deviation (SD) of the error, and the root mean square error. Bland-Altman limits of agreement (LoA) were computed, and equivalence tests were performed.ResultsForty patients were studied to compare the RVM and Respironics NM3 measurements. The mean difference (ie, bias) for MV was 1.8% with 95% LoA, defined as mean ± 1.96 SD, in the range of -12.1% to 15.7%. Similarly, the mean difference (ie, bias) for TV and RR was 1.2% (95% LoA, -11.0% to 13.5%) and 0.6% (95% LoA, -3.7% to 5.0%), respectively. The mean measurement precision of the RVM relative to the Respironics NM3 for MV, TV, and RR was 10.8%, 8.9%, and 8.4%, respectively. The mean measurement accuracy for MV, TV, and RR across patients was 11.0%, 9.7%, and 7.1%, respectively.ConclusionsThe data demonstrate that the RVM measures TV and MV in this cohort with an average relative error of 11% when using patient calibration and 16.9% without patient calibration. The average relative error of RR was 7.1%. The RVM provides accurate measurement of RR, TV, and MV in mechanically ventilated neonates and infants.Copyright © 2021 International Anesthesia Research Society.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.