• Thorax · Sep 1987

    Evaluation of the turbine pocket spirometer.

    • K A Gunawardena, K Houston, and A P Smith.
    • Riyadh Military Hospital, Saudi Arabia.
    • Thorax. 1987 Sep 1; 42 (9): 689-93.

    AbstractA compact electronic spirometer, the turbine pocket spirometer, which measures the FEV1, forced vital capacity (FVC), and peak expiratory flow (PEF) in a single expiration, was compared with the Vitalograph and the Wright peak flow meter in 99 subjects (FEV1 range 0.40-5.50 litres; FVC 0.58-6.48 l; PEF 40-650 l min-1). The mean differences between the machines were small--0.05 l for FEV1, 0.05 l for FVC, and 11.6 l min-1 for PEF, with the limits of agreement at +/- 0.25 l, +/- 0.48 l, and +/- 52.2 l min-1 respectively. The wide limits of agreement for the PEF comparison were probably because of the difference in the technique of blowing: a fast, long blow was used for the pocket spirometer and a short, sharp one for the Wright peak flow meter. The FEV1 and FVC showed a proportional bias of around 4-5% in favour of the Vitalograph. The repeatability coefficient for the pocket spirometer FEV1 was 0.18 l, for FVC 0.22 l, and for PEF 31 l min-1. These compared well with the repeatability coefficients of the Vitalograph and the Wright peak flow meter, which gave values of 0.18 l, 0.28 l, and 27 l min-1 respectively. At flow rates of over 600 l min-1 the resistance of the pocket spirometer marginally exceeded the American Thoracic Society recommendations. The machine is easy to operate and portable, and less expensive than the Vitalograph and Wright peak flow meter combined. It can be recommended for general use.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.