• Pacing Clin Electrophysiol · Oct 2014

    Safety and outcomes of magnetic resonance imaging in patients with abandoned pacemaker and defibrillator leads.

    • John V Higgins, Joseph J Gard, Seth H Sheldon, Raul E Espinosa, Christopher P Wood, Joel P Felmlee, Yong-Mei Cha, Samuel J Asirvatham, Connie Dalzell, Nancy Acker, Robert E Watson, and Paul A Friedman.
    • Department of Internal Medicine, Mayo Clinic, Rochester, Minnesota.
    • Pacing Clin Electrophysiol. 2014 Oct 1; 37 (10): 1284-90.

    IntroductionAbandoned cardiovascular implantable electronic device (CIED) leads remain a contraindication to magnetic resonance imaging (MRI) studies, largely due to in vitro data showing endocardial heating secondary to the radiofrequency field. We tested the hypothesis that abandoned CIED leads do not pose an increased risk of clinical harm for patients undergoing MRI.MethodsThis single-center retrospective study examined the outcomes of patients who had device generators removed before MRI, rendering the device leads abandoned. Information was gathered through chart review. Data collected included lead model, pacing threshold before MRI, anatomic region examined, threshold data after generator reimplantation, and clinical patient outcome.ResultsPatients (n = 19, 11 men and eight women) ranged in age from 19 to 85 at the time of MRI. There was a mean of 1.63 abandoned leads at the time of imaging; none of the leads were MRI conditional. Of the three implantable cardioverter defibrillator (ICD) leads, two of three were dual coil. Most (31/35) of the scans performed were of the central nervous system, including head and spinal imaging. There were no adverse events associated with MRI in any of these patients with abandoned leads within 7 days of the scan. No lead malfunctions or clinically significant change in pacing thresholds were noted with generator reimplantation.ConclusionThe use of MRI in patients with abandoned cardiac device leads appears feasible when performed under careful monitoring, with no adverse events, although the experience is small. MRI did not affect the function of leads that were subsequently reconnected to a cardiac device.©2014 Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…