- Sanjeeva P Kalva, Stephan Wicky, Arthur C Waltman, and Christos A Athanasoulis.
- Division of Cardiovascular Imaging & Intervention, Departments of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, Massachusetts, USA. skalva@partners.org
- J. Endovasc. Ther. 2006 Jun 1; 13 (3): 365-72.
PurposeTo evaluate the clinical safety and efficacy of the TrapEase vena cava filter in a 4-year single-center experience.MethodsThe clinical and imaging data of 751 patients (384 men; mean age 64 years, range 16-99) who had a TrapEase inferior vena cava (IVC) filter placed between January 1, 2001, and December 31, 2004, were reviewed retrospectively. More than a third of patients (297, 39.5%) presented with pulmonary embolism (PE), 188 (25.0%) had deep vein thrombosis (DVT), 40 (5.3%) had both PE and DVT, and the rest (226, 30.1%) had other symptoms. Indications for filter placement were contraindication to anticoagulation (461, 61.4%), complication of anticoagulation (42, 5.6%), failure of anticoagulation (39, 5.2%), and prophylaxis (209, 27.8%). Filters were placed in the infrarenal (n=738) or suprarenal (n=13) position through a femoral (n=729) or jugular vein (n=22) approach. Follow-up computed tomographic (CT) scans of the chest and abdomen were evaluated for recurrent PE and filter-related complications, respectively.ResultsThree (0.4%) patients developed groin hematoma. During a mean 295-day clinical follow-up (range 1-1677), 55 (7.5%) patients developed symptoms of PE, and 1 (0.1%) death was attributed to PE. Chest CT performed for various clinical indications in 219 patients at a mean 192 days (range 1-1346) showed PE in 15 (6.8%) patients; 10 were symptomatic and 5 asymptomatic, but there were no fatalities. Follow-up abdominal CT (n=270) at a mean 189 days (range 1-1415) showed fracture of filter components in 8 (3.0%), thrombus within the filter in 68 (25.2%), thrombus extending beyond the filter in 4 (1.5%), near total caval occlusion in 2 (0.7%), and no cases of migration.ConclusionThe TrapEase vena cava filter is effective in the prevention of pulmonary embolism, with minimal complications.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.