- Daniel B Brown, Dimitrios C Papadouris, Robert V Davis, Suresh Vedantham, and Thomas K Pilgram.
- Mallinckrodt Institute of Radiology, Washington University Medical Center, 510 South Kingshighway Boulevard, St. Louis, Missouri 63110, USA. brownda@mir.wustl.edu
- J Vasc Interv Radiol. 2005 Jan 1; 16 (1): 101-6.
PurposeTo determine the tolerance of 0.021-inch and 0.027-inch microcatheters to power injection in an in vitro flow model.Materials And MethodsTwenty-four microcatheters (0.021-inch, n = 13; 0.027-inch, n = 11) were injected with iothalamate meglumine through a flow model with use of a power injector and high-pressure tubing. Catheters used included Rebar (0.021-inch, n = 4; 0.027-inch, n = 4), Transit (0.021-inch, n = 3; 0.027-inch, n = 3), Renegade (0.021-inch, n = 4; 0.027-inch, n = 4), and Renegade STC-18 (0.021-inch, n = 2) models. Through the 0.021-inch microcatheters, 5-second injections were performed at an initial rate of 0.7 mL/sec. Injection rates were increased by 0.5 mL/sec and the process was repeated until the pressure approached 1,000 psi or catheter breakage occurred. A similar process was repeated for the 0.027-inch catheters starting at a rate of 3.4 mL/sec.ResultsThe 0.021-inch catheters were injected 303 times and the 0.027-inch catheters were injected 210 times. Three catheter failures occurred, with all breaks occurring at pressures greater than manufacturer recommendations. The 0.027-inch catheters as a group tolerated significantly higher injection rates than the 0.021-inch catheters. Of the 0.021-inch catheters, the STC-18 also provided superior maximum flow and volume compared with the Renegade catheter. The Rebar catheter tolerated significantly lower maximum injection rates and volumes than the other 0.027-inch catheters.ConclusionsThe majority of microcatheters can be power-injected in vitro at pressures far greater than manufacturer recommendations. When fractures occur, they are near the hub of the catheter. Significantly greater rates of injection are possible through 0.027-inch catheters.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.