• J Vasc Interv Radiol · Jan 2005

    Power injection of microcatheters: an in vitro comparison.

    • Daniel B Brown, Dimitrios C Papadouris, Robert V Davis, Suresh Vedantham, and Thomas K Pilgram.
    • Mallinckrodt Institute of Radiology, Washington University Medical Center, 510 South Kingshighway Boulevard, St. Louis, Missouri 63110, USA. brownda@mir.wustl.edu
    • J Vasc Interv Radiol. 2005 Jan 1; 16 (1): 101-6.

    PurposeTo determine the tolerance of 0.021-inch and 0.027-inch microcatheters to power injection in an in vitro flow model.Materials And MethodsTwenty-four microcatheters (0.021-inch, n = 13; 0.027-inch, n = 11) were injected with iothalamate meglumine through a flow model with use of a power injector and high-pressure tubing. Catheters used included Rebar (0.021-inch, n = 4; 0.027-inch, n = 4), Transit (0.021-inch, n = 3; 0.027-inch, n = 3), Renegade (0.021-inch, n = 4; 0.027-inch, n = 4), and Renegade STC-18 (0.021-inch, n = 2) models. Through the 0.021-inch microcatheters, 5-second injections were performed at an initial rate of 0.7 mL/sec. Injection rates were increased by 0.5 mL/sec and the process was repeated until the pressure approached 1,000 psi or catheter breakage occurred. A similar process was repeated for the 0.027-inch catheters starting at a rate of 3.4 mL/sec.ResultsThe 0.021-inch catheters were injected 303 times and the 0.027-inch catheters were injected 210 times. Three catheter failures occurred, with all breaks occurring at pressures greater than manufacturer recommendations. The 0.027-inch catheters as a group tolerated significantly higher injection rates than the 0.021-inch catheters. Of the 0.021-inch catheters, the STC-18 also provided superior maximum flow and volume compared with the Renegade catheter. The Rebar catheter tolerated significantly lower maximum injection rates and volumes than the other 0.027-inch catheters.ConclusionsThe majority of microcatheters can be power-injected in vitro at pressures far greater than manufacturer recommendations. When fractures occur, they are near the hub of the catheter. Significantly greater rates of injection are possible through 0.027-inch catheters.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…