• J Hand Ther · Oct 2018

    Measuring intrinsic hand strength in healthy adults: The accuracy intrarater and inter-rater reliability of the Rotterdam Intrinsic Hand Myometer.

    • Corey McGee.
    • Program in Occupational Therapy, University of Minnesota, Minneapolis, MN, USA. Electronic address: mcge0062@umn.edu.
    • J Hand Ther. 2018 Oct 1; 31 (4): 530-537.

    Study DesignClinical measurement study.IntroductionMeasuring the isometric strength generated during isolated hand joint motions is a challenging feat. The Rotterdam Intrinsic Hand Myometer (RIHM; med.engineers, Rotterdam, Netherlands) permits measurement of isolated movements of the hand. To date, there is limited evidence on the inter-rater reliability and limited adult normative data of RIHM. Given that multiple raters, often with varying degrees of experience, are needed to collect normative data, inter-rater reliability testing and a comparison of novice and experienced raters are needed.Purposes Of The StudyThe purposes of this study were to test the accuracy, intrarater reliability, and inter-rater reliability of the RIHM in healthy-handed adults.MethodsRIHM accuracy was tested through use of precision class F weights. Adults 18 years or older without upper limb dysfunction were recruited. Each participant was tested by 4 raters, 3 occupational therapy graduate students, and an experienced certified hand therapist, through use of a calibrated RIHM. Five strength measures were tested bilaterally (ie, thumb carpometacarpal palmar abduction, index finger metacarpophalangeal [MP] abduction, index finger MP flexion, thumb MP flexion, and small finger MP abduction) 3 times per a standardized protocol. Statistical methods were used to test accuracy, inter-rater reliability, and intrarater/response stability.ResultsThe accuracy of RIHM device error was 5% or less. Reliability testing included the participation of 19 women and 10 men (n = 29). All raters were in excellent agreement across all muscles (intraclass correlation coefficient, ≥0.81). Low standard error of measurement values of ≤8.3 N (1.9 lb) across raters were found. The response stability and/or intrarater reliability of the novice and certified hand therapist raters were not statistically different.DiscussionThe RIHM has an acceptable instrument error; the RIHM and its standardized procedure have excellent inter-rater reliability and response stability when testing those without hand limitations; and the response stability and/or intrarater reliability of expert and novice raters were consistent.ConclusionsThe use of the RIHM is justified when multiple raters of varying expertise collect normative data or conduct cohort studies on persons with healthy hands. Future research is warranted.Level Of EvidenceNot applicable.Copyright © 2017 Hanley & Belfus. Published by Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…