-
ANZ journal of surgery · Jan 2011
Influence of screw diameter and number on reduction loss after plating of distal radius fractures.
- Herwig Drobetz, Michael Schueller, Elmar Karl Tschegg, Clare Heal, Heinz Redl, and Reinhold Muller.
- Department of Orthopaedic Surgery, Mackay Base Hospital, Mackay, Queensland, Australia. drobo@aon.at
- ANZ J Surg. 2011 Jan 1; 81 (1-2): 46-51.
BackgroundThe current options for plate-screw combinations in volar locking distal radius plates used for the treatment of distal radius fractures are either plates with a single distal screw row or plates with multiple distal screw rows. Additionally, the screws themselves may have either fixed angle locking or polyaxial locking mechanisms. To date, there is no evidence or consensus regarding the optimal plate-screw combination. The aim of this study was to assess the biomechanical behaviour of different plate-screw combinations with respect to total distal screw number, number of distal screw rows and screw projection surface area of the most distal row.MethodsBiomechanical study to assess six different plating configurations in five different volar locking plate models in a Sawbone distal radius fracture model. The specimens were loaded with 800 Newton loads for 2.000 cycles at 1 Hz. After cyclic loading, load-to-failure testing was performed.Results With cyclical testing, there was a significant and positive correlation between rigidity and a greater projection area of the most distal screws. Dorsal tilting was significantly more pronounced in plate models with a lesser projection area of the most distal screws and a smaller number of distal screws. With load-to-failure testing, there was a significant increase in rigidity with increasing screw projection area of the most distal row and total number of distal screws.ConclusionsAdditional distal screw rows in volar locking distal radius plates might not add substantially to resistance against loss of reduction in the post-operative period.© 2010 The Authors. ANZ Journal of Surgery © 2010 Royal Australasian College of Surgeons.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.