    The variability of results between point-of-care testing glucose meters and the central laboratory analyzer.

    • Adil I Khan, Yolanda Vasquez, Jacquelyn Gray, Frank H Wians, and Martin H Kroll.
    • Department of Pathology, University of Texas Southwestern Medical Center, Dallas, USA.
    • Arch. Pathol. Lab. Med. 2006 Oct 1; 130 (10): 1527-32.

    ContextPoint-of-care testing glucose meters are strongly recommended in the management of diabetes and are increasingly being used for making therapeutically important decisions. Thus, it is essential that their results correlate well with those of laboratory analyzers.ObjectivesTo test the reliability of point-of-care testing glucose meters.DesignTwo studies were performed: (1), an in-house study comparing accuracy of point-of-care testing glucose meters with a reference analyzer using fresh whole blood specimens (2), a real-time comparison of (a) 2 successive glucose meter readings and (b) glucose meter reading to central laboratory analyzer reading.Setting(1), Seven glucose meters from 4 manufacturers were compared with the Yellow Springs YSI 2300 blood glucose analyzer using whole blood without preservative. (2), (a) Whole blood samples were read within 5 minutes of each other using Accu-Chek meters and (b) between a glucose meter and a Hitachi laboratory analyzer.Results(1) Within the Accu-Chek group of glucose meters, fresh, preservative-free whole blood samples showed the lowest bias. (2) At the hypoglycemic level, successive glucose meter readings agreed well, but there was considerable disagreement between glucose meter and central laboratory values. Because laboratory analyzers are of proven accuracy, they are used as the reference. In the glucose meter-central laboratory analyzer correlation, for both hypoglycemic and hyperglycemic values, readings in which the differences were greater than 10% occurred more than 61% of the time. In the hypoglycemic range, differences greater than 20% occurred 57% of the time.ConclusionsOne should scrutinize point-of-care testing glucose meter readings at the hypoglycemic and hyperglycemic levels and whenever possible to corroborate these clinical results with central laboratory analyzers.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…