• Diabetes care · Dec 1996

    Comparative Study

    Accuracy of home blood glucose meters during hypoglycemia.

    • Z Trajanoski, G A Brunner, R J Gfrerer, P Wach, and T R Pieber.
    • Department of Biophysics, Graz University of Technology, Austria. trajanoski@ibmt.tu-graz.ac.at
    • Diabetes Care. 1996 Dec 1; 19 (12): 1412-5.

    ObjectiveTo evaluate the accuracy of home blood glucose meters during hypoglycemia.MethodsSix blood glucose meters-One Touch II (LifeScan, Milpitas, CA), Companion II (Medisense, Cambridge, U.K.), Reflolux (Boehringer Mannheim, Mannheim, Germany), Accutrend (Boehringer Mannheim), Elite (Bayer, Munich, Germany), and HemoCue (HemoCue, Angelholm, Sweden)-were compared with a reference method (Beckman Glucose Analyzer 2). Glucose concentrations from arterialized venous blood samples were measured using all glucose meters (whole blood) and the reference method (plasma) during hypoglycemic-hyperinsulinemic clamps in 15 subjects.ResultsIn total, 663 blood glucose monitor readings and 119 reference values ranging from 2.28 to 3.89 mmol/l were analyzed. The correlation coefficients and the percentage of measurements within 20% and outside 40% of the reference values for each glucose meter were as follows: One Touch II: 0.91, 99.2% and 0%; Companion II: 0.81, 88.2% and 2.5%; Reflolux: 0.78, 85.0% and 0.9%; Accutrend: 0.88, 46.0% and 6.6%; Elite: 0.78, 75.6% and 4.2%; and HemoCue: 0.93, 96.6% and 0% (P < 0.001).ConclusionsThere were substantial differences between the blood glucose meters during hypoglycemia, and none of the devices met the latest criteria recommended by the American Diabetes Association.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.