• Can J Anaesth · Sep 2021

    Dimensional compatibility and limitations of tracheal intubation through supraglottic airway devices: a mannequin-based in vitro study.

    • Berthold Moser, Michael Kemper, Maren Kleine-Brueggeney, Lukas Gasteiger, and Markus Weiss.
    • Department of Anaesthesia and Intensive Care, Medical University of Innsbruck, Anichstraße 35, 6020, Innsbruck, Austria.
    • Can J Anaesth. 2021 Sep 1; 68 (9): 133713481337-1348.

    BackgroundFlexible bronchoscope-guided tracheal intubation through supraglottic airway devices (SGAs) is a well-established element of difficult intubation algorithms. Success can be limited by dimensional incompatibilities between tracheal tubes (TTs) and SGAs.MethodsIn this in vitro study, we tested the feasibility of TT passage through SGAs, removal of SGAs over TTs, and the ability to guide the flexible bronchoscope with 13 TT brands (internal diameter, 6.5-8.0 mm) and ten different SGAs (#4 and #5) in an intubation mannequin.ResultsWe tested 1,040 combinations of SGAs and TTs. Tracheal tube passage failed in 155 (30%) combinations of the five tested first-generation SGAs (117 [46%] with SGA #4, 38 [15%] with SGA #5) and in three (0.6%) combinations of the five tested second-generation SGAs (two [0.8%] with SGA #4 and one [0.4%] with SGA #5). The reason for failed passage of a TT through a first-generation SGA consistently was a too-narrow SGA connector. Removal of the SGA over the TT in the 882 remaining combinations was impossible for all sizes of reinforced TTs, except the Parker Reinforced TT, and was possible for all non-reinforced TTs. Only one combination with SGA #4 and 84 combinations with SGA #5 were not ideal to adequately guide the flexible bronchoscope.ConclusionClinically relevant combinations of adult-size TTs and SGAs can be incompatible, rendering flexible bronchoscope-guided tracheal intubation through an SGA impossible. Additional limitations exist regarding removal of the SGA and maneuverability of the flexible bronchoscope.© 2021. The Author(s).

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…