Randomized Controlled Trial Comparative Study
Coronary sinus electrode does not reduce atrial defibrillation thresholds.
- Eric J Rashba, Stephen R Shorofsky, Avram Scheiner, Robert W Peters, Carol Ma, and Michael R Gold.
- Division of Cardiology, University of Maryland School of Medicine, Baltimore, 21201, USA. erashba@medicine.umaryland.edu
- Heart Rhythm. 2006 Jun 1; 3 (6): 647-52.
BackgroundAtrial defibrillation can be achieved with a conventional dual-coil, active pectoral implantable cardioverter-defibrillator (ICD) lead system. Shocking vectors that incorporate an additional electrode in the CS have been used, but it is unclear if they improve atrial DFTs.ObjectiveThe objective of this prospective, randomized study was to determine if a coronary sinus (CS) electrode reduces atrial defibrillation thresholds (DFTs).MethodsThis was a prospective study of 36 patients undergoing initial ICD implant for standard indications. A defibrillation lead with superior vena cava (SVC) and right ventricular (RV) shocking coils was implanted in the RV. An active can emulator (Can) was placed in a pre-pectoral pocket. A lead with a 4 cm long shocking coil was placed in the CS. Atrial DFTs were determined in the following 3 shocking configurations in each patient, with the order of testing randomized: RV --> SVC + Can (Ventricular Triad), distal CS --> SVC + Can (Distal Atrial Triad), and proximal CS --> SVC + Can (Proximal Atrial Triad).ResultsThe Proximal and Distal Atrial Triad configurations were both associated with significant reductions in peak current (p < 0.01), but this effect was offset by significant increases in shock impedance (p < 0.01), resulting in no net change in the peak voltage or DFT energy in comparison to the Ventricular Triad configuration (Ventricular Triad: 4.9 +/- 6.6 J, Proximal Atrial Triad: 3.3 +/- 4.1J, Distal Atrial Triad: 4.4 +/- 6.7 J, p > 0.2).ConclusionShocking vectors that incorporate a CS coil do not significantly improve atrial defibrillation efficacy. Since the Ventricular Triad shocking pathway provides reliable atrial and ventricular defibrillation, this configuration should be preferred for combined atrial and ventricular ICDs.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.