• Intensive care medicine · Jan 2001

    In vitro validation of a metabolic monitor for gas exchange measurements in ventilated neonates.

    • M Behrends, M Kernbach, A Bräuer, U Braun, J Peters, and W Weyland.
    • Abteilung für Anästhesiologie und Intensivmedizin, Univerversitätsklinkum Essen, Germany. matthias.behrends@uni-essen.de
    • Intensive Care Med. 2001 Jan 1; 27 (1): 228-35.

    ObjectiveTo evaluate the Datex Deltatrac II for measurements in neonates requiring mechanical ventilation.DesignProspective laboratory evaluation, using a ventilated lung model and gas injection. During simulation of 79 neonatal respiratory settings, assessment of oxygen consumption (VO2), carbon dioxide production (VCO2) and respiratory quotient (RQ) was compared to a reference method (mass spectrometry, wet gas spirometry) using the statistical method of Bland and Altman.InterventionsRespiratory variables, which may influence the accuracy and precision of gas exchange measurements, were varied within the following ranges: inspired oxygen fraction (FIO2): 0.21-0.8, expired carbon dioxide fraction (FECO2) and inspiratory-expiratory oxygen fraction (DFO2): 0.0032-0.0256, expiratory flow rate: 1.0-2.5 l/min, inspiratory pressure: 10-55 mbar, respiratory rate 25-60/min, constant RQ of 1. This resulted in 79 tests with VCO2 and VO2 ranging from 8-64 ml/min.Measurements And ResultsThe coefficient of repeatability for ten single subsequent Deltatrac measurements was 8.09 ml/min for VO2 and 9.17 ml/min for VCO2 compared to 2.02 ml/min and 0.90 ml/min for VO2 and VCO2 with repeated reference measurements. The coefficient of repeatability of the Deltatrac measurements improved considerably when means of subsequent 5 min intervals were compared: 0.68 ml/min for VO2 and 0.28 ml/ min for VCO2. The difference between the two methods (Deltatrac-reference) was -3.8 % (2 s: 11.4%) for VO2, 13.2% (2s: 7.9%) for VCO2 and 17.6% (2s: 16.7%) for RQ. The agreement between methods deteriorated with smaller (FECO2) or DFO2 and increasing FIO2.ConclusionsConsidering limits of agreement of less than +/- 20% as clinically acceptable, results for VO2 assessment indicate acceptable accuracy and precision whereas VCO2 and RQ assessments exceed this limit. Limited accuracy and precision result from detection of CO2 following dilution of expiratory gases and increased sensitivity to error propagation by Haldane equations due to the small differences between inspiratory and expiratory gas fractions.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…