- Marta Delsoglio, Yves Marc Dupertuis, Taku Oshima, Mart van der Plas, and Claude Pichard.
- Clinical Nutrition, Geneva University Hospital, Geneva, Switzerland. Electronic address: Marta.DELSOGLIO@hcuge.ch.
- Clin Nutr. 2020 Jun 1; 39 (6): 1927-1934.
Background & AimsIndirect calorimetry (IC) is the only way to measure in real time energy expenditure (EE) and to optimize nutrition support in acutely and chronically ill patients. Unfortunately, most of the commercially available indirect calorimeters are rather complex to use, expensive and poorly accurate and precise. Therefore, an innovative device (Q-NRG®, COSMED, Rome, Italy) that matches clinicians' needs has been developed as part of the multicenter ICALIC study supported by the two academic societies ESPEN and ESICM. The aim of this study was to evaluate the accuracy and intra- and inter-unit precision of this new device in canopy dilution mode in vitro and in spontaneously breathing adults.MethodsAccuracy and precision of oxygen consumption (VO2) and carbon dioxide production (VCO2) measurements were evaluated in vitro and in 15 spontaneously breathing healthy adults by interchanging three Q-NRG® units in a random order. In vitro validation was performed by gas exchange simulation using high-precision gas mixture and mass flow controller. Accuracy was calculated as error of measured values against expected ones based on volume of gas infused. Respiratory coefficient (RQ) accuracy was furthermore assessed using the ethanol-burning test. To evaluate the intra- and inter-unit precisions, the coefficient of variation (CV% = SD/Mean*100) was calculated, respectively, from the mean ± SD or the mean ± SD of the three mean values of VO2, VCO2, RQ and EE measured by each Q-NRG® units. In vivo accuracy measurement of the Q-NRG® was assessed by simultaneous comparison with mass spectrometry (MS) gas analysis, using Bland-Altman plot, Pearson correlation and paired t-test (significance level of p = 0.05).ResultsIn vitro evaluation of the Q-NRG® accuracy showed measurement errors <1% for VO2, VCO2 and EE and <1.5% for RQ. Evaluation of the intra- and inter-unit precision showed CV% ≤1% for VO2 and EE and CV% ≤1.5% for VCO2 and RQ measurements, except for one Q-NRG® unit where CV% was 2.3% for VO2 and 3% for RQ. Very good inter-unit precision was confirmed in vivo with CV% equal to 2.4%, 3%, 2.8% and 2.3% for VO2, VCCO2, RQ and EE, respectively. Comparison with MS showed correlation of 0.997, 0.987, 0.913 and 0.997 for VO2, VCO2, RQ and EE respectively (p ≤ 0.05). Mean deviation of paired differences was 1.6 ± 1.4% for VO2, -1.5 ± 2.5% for VCO2, -3.1 ± 2.6% for RQ and 0.9 ± 1.4% for EE.ConclusionBoth in vitro and in vivo measurements of VO2, VCO2, RQ and EE on three Q-NRG® units showed minimal differences compared to expected values and MS and very low intra- and inter-unit variability. These results confirm the very good accuracy and precision of the Q-NRG® indirect calorimeter in canopy dilution mode in spontaneously breathing adults.Copyright © 2019 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.