- Amin Joukar, Ali Kiapour, Anoli Shah, Ardalan Seyed Vosoughi, and Vijay K Goel.
- Departments of Bioengineering and Orthopaedic Surgery, Engineering Center for Orthopaedic Research Excellence (E-CORE), Colleges of Engineering and Medicine, University of Toledo, 5046 NI, MS 303, Toledo, OH, 43606, USA.
- Eur Spine J. 2021 Aug 1; 30 (8): 2351-2359.
PurposeThis study's objective was to assess biomechanical parameters across fused and contralateral sacroiliac joints (SIJs) and implants during all spinal motions for both sexes. Various SIJ implant devices on the market are used in minimally invasive surgeries. These implants are placed across the joint using different surgical approaches. The biomechanical effects of fusion surgical techniques in males and females have not been studied.MethodsThe validated finite element models of a male, and a female spine-pelvis-femur were unilaterally instrumented across the SIJ using three screws for two SIJ implants, half threaded and fully threaded screws placed laterally and posteriorly to the joint, respectively.ResultsMotion and peak stress data at the SIJs showed that the female model exhibited lower stresses and higher reduction in motion at the contralateral SIJ in all motions than the male model predictions with 84% and 71% reductions in motion and stresses across the SIJ.ConclusionImplants exhibited higher stresses in the female model compared to the male model. However, chances of SIJ implant failure in the female patients are still minimal, based on the calculated factor of safety which is still very high. Both lateral and posterior surgical approaches were effective in both sexes; however, the lateral approach may provide a better biomechanical response, especially for females. Moreover, implant design characteristics did not make a difference in the implants' biomechanical performance. SIJ stabilization was primarily provided by the implants which were the farthest from the sacrum rotation center.© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.