• Interact Cardiovasc Thorac Surg · Jun 2016

    Comparative Study Clinical Trial

    Conventional versus rapid-deployment aortic valve replacement: a single-centre comparison between the Edwards Magna valve and its rapid-deployment successor.

    • Martin Andreas, Stephanie Wallner, Andreas Habertheuer, Claus Rath, Martin Schauperl, Thomas Binder, Dietrich Beitzke, Raphael Rosenhek, Christian Loewe, Dominik Wiedemann, Alfred Kocher, and Guenther Laufer.
    • Department of Surgery, Division of Cardiac Surgery, Medical University of Vienna, Vienna, Austria martin.andreas@meduniwien.ac.at.
    • Interact Cardiovasc Thorac Surg. 2016 Jun 1; 22 (6): 799-805.

    ObjectivesSutureless and rapid-deployment valves were recently introduced into clinical practice. The Edwards INTUITY valve system is a combination of the Edwards Magna pericardial valve and a subvalvular stent-frame to enable rapid deployment. We performed a parallel cohort study for comparison of the two valve types.MethodsAll patients receiving either an Edwards Magna Ease valve or an Edwards INTUITY valve system due to aortic stenosis from May 2010 until July 2014 were included. Patients undergoing bypass surgery, an additional valve procedure, atrial ablation surgery or replacement of the ascending aorta were excluded. Preoperative characteristics, operative specifications, survival, valve-related adverse events and transvalvulvar gradients were compared.ResultsOne hundred sixteen patients underwent rapid-deployment aortic valve replacement [mean age 75 years (SD: 8); 62% female] and 132 patients underwent conventional aortic valve replacement [70 years (SD: 9); 31% female; P < 0.001]. Conventional valve patients were taller and heavier. The mean EuroSCORE II was 3.1% (SD: 2.7) and 4.4% (SD: 6.0) for rapid-deployment and conventional valve patients, respectively (P = 0.085). The mean implanted valve size was higher in the conventional group [23.2 mm (SD: 2.0) vs 22.5 mm (SD: 2.2); P = 0.007], but postoperative transvalvular mean gradients were comparable [15 mmHg (SD: 6) vs 14 mmHg (SD: 5); P = 0.457]. A subgroup analysis of the most common valve sizes (21 and 23 mm; implanted in 63% of patients) revealed significantly reduced mean postoperative transvalvular gradients in the rapid-deployment group [14 mmHg (SD: 4) vs 16 mmHg (SD: 5); P = 0.025]. A significantly higher percentage received minimally invasive procedures in the rapid-deployment group (59 vs 39%; P < 0.001). The 1- and 3-year survival rate was 96 and 90% in the rapid-deployment group and 95 and 89% in the conventional group (P = 0.521), respectively. Valve-related pacemaker implantations were more common in the rapid-deployment group (9 vs 2%; P = 0.014) and postoperative stroke was more common in the conventional group (1.6 vs 0% per patient year; P = 0.044).ConclusionsWe conclude that this rapid-deployment valve probably facilitates minimally invasive surgery. Furthermore, a subgroup analysis showed reduced transvalvular gradients in smaller valve sizes compared with the conventionally implanted valve of the same type. The favourable haemodynamic profile and the potentially different spectrum of valve-related adverse events should be addressed in further clinical trials.© The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…