- G Etienne Cramer, Peter C Kievit, Marc A Brouwer, Marinus H de Keijzer, Hans E Luijten, and Freek W A Verheugt.
- Heartcentre, Radboud University Nijmegen Medical Centre, The Netherlands. g.cramer@cardio.umcn.nl
- Clin. Chim. Acta. 2007 Jun 1; 381 (2): 164-6.
ObjectivesThis study was designed to test the usefulness of a bedside assay as compared to a laboratory method of troponin testing to predict adverse cardiac outcome of chest pain patients.MethodsWe studied 358 ER visits of patients suspected of a non ST-elevation acute coronary syndrome. cTnI (Immulite, DPC) on a lab analyser and cTnT (Cardiac Reader, Roche) at bedside were measured at baseline. The between-assay level of concordance, reporting turnaround times and clinical outcomes during 180 days of follow-up were assessed. Death and myocardial infarction were then evaluated according to troponin result, either concordant negative, discordant or concordant positive.ResultsDiscordance occurred in 11.4% (41/358) of cases. The proportion of patients with a positive cTnI and negative cTnT result (8.9%) versus the reverse (2.5%) differed significantly (p<0.001). The median time gained using the rapid test was 72 min. The rate of death and/or MI was 25% (10/40) among patients with discordant results as compared to 7.5% (17/228) with a concordant negative result (p<0.001). All patients from the discordant group with an event had a positive cTnI result, while cTnT was negative.ConclusionPatients with a discordant reading were at high risk of adverse cardiac outcome, which was only identified by the laboratory cTnI assay. Markedly, the use of the rapid assay saved time at the expense of clinical sensitivity.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.