Comparative StudyPilot study of pulmonary arterial branch sealing using energy devices in an ex vivo model.
- Moishe Liberman, Mohamed Khereba, Eric Goudie, Jordan Kazakov, Vicky Thiffault, Edwin Lafontaine, and Pasquale Ferraro.
- CETOC - CHUM Endoscopic Tracheobronchial and Oesophageal Center, Division of Thoracic Surgery, Department of Surgery, University of Montréal, Montréal, Québec, Canada. Electronic address: moishe.liberman@umontreal.ca.
- J. Thorac. Cardiovasc. Surg.. 2014 Dec 1;148(6):3219-23.
ObjectiveVascular endostaplers are bulky and can be dangerous when dividing small pulmonary arterial (PA) branch vessels during video-assisted thoracoscopic lobectomy. We aimed to evaluate and compare the immediate efficacy of modern energy sealing devices in an ex vivo PA sealing model.MethodsPatients undergoing anatomical lung resection or lung transplantation were recruited for a prospective cohort pilot study. Four devices were evaluated: Harmonic Ace (Ethicon, Cincinnati, Ohio), Thunderbeat (Olympus, Tokyo, Japan), LigaSure (Covidien, Boulder, Colo), and Enseal (Ethicon; Cincinnati, Ohio). After anatomical lung resection, the PA branches were dissected in vitro. Sealing was then performed with 1 of the sealing devices, the vessel was slowly pressurized, and the bursting pressure was recorded.ResultsForty-nine PA branches were sealed in 14 patients. The mean PA branch diameter was 7.4 mm (1.8-14.5 mm). Ten patients had normal PA pressure and 3 had PA hypertension. The mean bursting pressure in each was as follows: Harmonic Ace group, 415.5 mm Hg (137.1-1388.4 mm Hg), Thunderbeat group, 875 mm Hg (237.1-2871.3 mm Hg); LigaSure group, 214.7 mm Hg (0-579.6 mm Hg); Enseal group, 133.7 mm Hg (0-315.38 mm Hg). There were 2 complete sealing failures: LigaSure (diameter 6.78 mm) and Enseal (diameter 8.3 mm).ConclusionsIn this pilot study to examine energy sealing of PA branches in a simulated ex vivo model, vascular sealing using energy was effective and was able to sustain high intraluminal bursting pressures. Further research is needed to determine the in vivo and long-term safety of PA branch energy sealing.Copyright © 2014 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.