• Medicine · Jun 2021

    Observational Study

    Blood pressure measurement during cesarean delivery: Evaluation of a beat-to-beat noninvasive device (NexfinTM).

    • Mathieu Bobet, Jona Joachim, Etienne Gayat, Agnès Bonnet, Kerstin Sievert, Carole Barnichon, Marc Fischler, and Morgan Le Guen.
    • Department of Anesthesiology, Hôpital Foch, Suresnes, France and University Versailles Saint-Quentin en Yvelines, Montigny-Le-Bretonneux.
    • Medicine (Baltimore). 2021 Jun 4; 100 (22): e26129e26129.

    AbstractEarly detection of arterial hypotension during cesarean delivery under spinal anesthesia is important. This study aims to compare the validity of NexfinTM as beat-to-beat noninvasive blood pressure monitoring with conventional intermittent oscillometric measurement of blood pressure during elective cesarean delivery.This open prospective observational bicentric study was performed between January 2013 and December 2015. We simultaneously recorded arterial blood pressure with both techniques in pregnant women undergoing elective cesarean delivery under spinal anesthesia. The primary outcome was a Bland-Altman analysis of systolic blood pressure measurement comparing NexfinTM and a conventional method. The secondary outcomes were the time to detect the first relevant hypotensive episode and the comparison of both devices using a four-quadrant graph.One hundred and seventy-four parturients completed the study, and 2640 pairs of systolic blood pressure measurements were analyzed. Bias was -10 mmHg with upper and lower limits of agreement of -61 and +41 mmHg. In 73.9% of the cases, the two techniques provided the same information (normotension or hypotension), but the conventional method missed 20.8% of measurements, with NexfinTM detecting 16.2% more hypotensive measurements. The median [25-75 percentiles] duration to detect the first hypotensive measurement was 331 [206-480] seconds for NexfinTM and 440 [300-500] s for intermittent oscillometry (P < .001).The agreement between NexfinTM and an intermittent method for the measurement of systolic blood pressure was not in an acceptable range during cesarean delivery, although NexfinTM may detect hypotension earlier than the standard method.Trial registration: Clinicaltrials.gov identifier: NCT01732133; November 22, 2012.Copyright © 2021 the Author(s). Published by Wolters Kluwer Health, Inc.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…