• J Electrocardiol · Mar 2018

    Evaluation of ECG algorithms designed to improve detect of transient myocardial ischemia to minimize false alarms in patients with suspected acute coronary syndrome.

    • Michele M Pelter, Yuan Xu, Richard Fidler, Ran Xiao, David W Mortara, and Hu Xiao.
    • Department of Physiological Nursing, University of California San Francisco, San Francisco, CA, United States. Electronic address: michele.pelter@ucsf.edu.
    • J Electrocardiol. 2018 Mar 1; 51 (2): 288-295.

    BackgroundPatients hospitalized for suspected acute coronary syndrome (ACS) are at risk for transient myocardial ischemia. During the "rule-out" phase, continuous ECG ST-segment monitoring can identify transient myocardial ischemia, even when asymptomatic. However, current ST-segment monitoring software is vastly underutilized due to false positive alarms, with resultant alarm fatigue. Current ST algorithms may contribute to alarm fatigue because; (1) they are not designed with a delay (minutes), rather alarm to brief spikes (i.e., turning, heart rate changes), and (2) alarm to changes in a single ECG lead, rather than contiguous leads.PurposeThis study was designed to determine sensitivity, and specificity, of ST algorithms when accounting for; ST magnitude (100μV vs 200μV), duration, and changes in contiguous ECG leads (i.e., aVL, I, - aVR, II, aVF, III; V1, V2, V3, V4, V5, V6, V6, I).MethodsThis was a secondary analysis from the COMPARE Study, which assessed occurrence rates for transient myocardial ischemia in hospitalized patients with suspected ACS using 12-lead Holter. Transient myocardial ischemia was identified from Holter using >100μV ST-segment ↑ or ↓, in >1 ECG lead, >1min. Algorithms tested against Holter transient myocardial ischemia were done using the University of California San Francisco (UCSF) ECG algorithm and included: (1)100μV vs 200μV any lead during a 5-min ST average; (2)100μV vs 200μV any lead >5min, (3) 100μV vs 200μV any lead during a 5-min ST average in contiguous leads, and (4) 100μV vs 200μV>5min in contiguous leads (Table below).ResultsIn 361 patients; mean age 63+12years, 63% male, 56% prior CAD, 43 (11%) had transient myocardial ischemia. Of the 43 patients with transient myocardial ischemia, 17 (40%) had ST-segment elevation events, and 26 (60%) ST-segment depression events. A higher proportion of patients with ST segment depression has missed ischemic events. Table shows sensitivity and specificity for the four algorithms tested.ConclusionsSensitivity was highly variable, due to the ST threshold selected, with the 100μV measurement point being superior to the 200μV amplitude threshold. Of all the algorithms tested, there was moderate sensitivity and specificity (70% and 68%) using the 100μV ST-segment threshold, integrated ST-segment changes in contiguous leads during a 5-min average.Copyright © 2017 Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.