-
- Kenneth L Urish, Brian R Hamlin, Anton Y Plakseychuk, Timothy J Levison, Genymphas B Higgs, Steven M Kurtz, and Anthony M DiGioia.
- Bone and Joint Center, Magee Womens Hospital of the University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania; Department of Orthopaedic Surgery, Department of Bioengineering, and Clinical and Translational Science Institute, University of Pittsburgh; Department of Biomedical Engineering, Carnegie Mellon University; Pittsburgh, Pennsylvania.
- J Arthroplasty. 2017 Sep 1; 32 (9): 2857-2863.
BackgroundGross trunnion failure (GTF) is a rare complication in total hip arthroplasty (THA) reported across a range of manufacturers. Specific lots of the Stryker low friction ion treatment (LFIT) anatomic cobalt chromium alloy (CoCr) V40 femoral head were recalled in August 2016. In part, the recall was based out of concerns for disassociation of the femoral head from the stem and GTF.MethodsWe report on 28 patients (30 implants) with either GTF (n = 18) or head-neck taper corrosion (n = 12) of the LFIT CoCr femoral head and the Accolade titanium-molybdenum-zirconium-iron alloy femoral stems. All these cases were associated with adverse local tissue reactions requiring revision of the THA.ResultsIn our series, a conservative estimate of the incidence of failure was 4.7% (n = 636 total implanted) at 8.0 ± 1.4 years from the index procedure. Failures were associated with a high-offset 127° femoral stem neck angle and increased neck lengths; 43.3% (13 of 30) of the observed failures included implant sizes outside the voluntary recall (27.8% [5 of 18] of the GTF and 75.0% [8 of 12] of the taper corrosion cases). Serum cobalt and chromium levels were elevated (cobalt: 8.4 ± 7.0 μg/mL; chromium: 3.4 ± 3.3 μ/L; cobalt/chromium ratio: 3.7). The metal artifact reduction sequence magnetic resonance imaging demonstrated large cystic fluid collections typical with adverse local tissue reactions. During revision, a pseudotumor was observed in all cases. Pathology suggested a chronic inflammatory response. Impending GTF could be diagnosed based on aspiration of black synovial fluid and an oblique femoral head as compared with the neck taper on radiographs.ConclusionIn our series of the recalled LFIT CoCr femoral head, the risk of impending GTF or head-neck taper corrosion should be considered as a potential diagnosis in a painful LFIT femoral head and Accolade titanium-molybdenum-zirconium-iron alloy THA with unknown etiology. Almost half of the failures we observed included sizes outside of the voluntary recall.Copyright © 2017 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.