• Interact Cardiovasc Thorac Surg · Jul 2017

    Case Reports

    Severe aortic regurgitation after implantation of a sutureless valve prosthesis using an automatic knot fastener device.

    • Robert Balan, Christian Mogilansky, Axel-Lars Larsen, and Parwis Massoudy.
    • Department of Cardiac Surgery, Klinikum Passau, Passau, Germany.
    • Interact Cardiovasc Thorac Surg. 2017 Jul 1; 25 (1): 153-154.

    AbstractWe report the case of severe aortic regurgitation 8 months after implantation of a 25-mm sutureless pericardial aortic valve prosthesis. On echocardiography, the regurgitation was suspected to be paravalvular. The sutureless prosthesis had been implanted using an automatic knot fastener device, which renders the suture tails less pliable because of the metal clip that is crimped around the suture. The patient was reoperated, a paravalvular leak was not observed. The sutureless prosthesis was explanted and a conventional biologic valve prosthesis was implanted instead. On examination of the explanted valve prosthesis, a perforation was observed in one of the leaflets. The leaflet perforation was in alignment with one of the knots produced by the automatic knot fastener. Obviously, the leaflet had hit the knot repeatedly which had caused the perforation. We conclude that knots produced by an automatic fastener device have the potential to cause leaflet perforation.© The Author 2017. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…