• World Neurosurg · Feb 2016

    Intraoperative Fluoroscopy for Ventriculoperitoneal Shunt Placement.

    • Daniel Coluccia, Javier Anon, Frederic Rossi, Serge Marbacher, Javier Fandino, and Sven Berkmann.
    • Department of Neurosurgery, Kantonsspital Aarau, Aarau, Switzerland. Electronic address: daniel.coluccia@ksa.ch.
    • World Neurosurg. 2016 Feb 1; 86: 71-8.

    ObjectiveCatheter malpositioning is one of the most frequent causes of ventriculoperitoneal shunt dysfunction and revision surgery. Most intraoperative tools used to improve the accuracy of catheter insertion are time consuming and expensive or do not display the final position. We evaluate the usefulness of intraoperative fluoroscopy to decrease catheter malpositioning, and define radiological landmarks to identify the correct localization.MethodsA total of 104 patients undergoing ventriculoperitoneal shunt placement were analyzed for shunt position, revision surgery and outcome. The results for patients operated on using intraoperative biplanar fluoroscopic assessment of catheter location (X-ray group, n = 57) were compared with a control group operated without intraoperative radiography (control, n = 47). In order to generate a surgical reference map for intraoperative validation of shunt location, different ventricular system landmarks were defined on three-dimensional computed tomography reconstructions of hydrocephalic patients (n = 60) and exported to a two-dimensional layer of the skull.ResultsThe use of intraoperative X-ray imaging correlated with a significant increase of optimal catheter positions (X-ray group, n = 45, 79%; control group, n = 23, 49%; P = 0.0018). The sensitivity and positive predictive value for estimating an optimal shunt catheter position on biplanar imaging was 96% (95% confidence interval, 87%-99%). The specificity and negative predictive value were both 92% (95% confidence interval, 78%-98%).ConclusionsIntraoperative fluoroscopy is easy to perform and is a reliable method to assess correct catheter positioning. Based on its predictive value, corrections of malpositioned ventricular catheters can be performed during the same procedure. The use of intraoperative fluoroscopy decreases early surgical revisions in ventriculoperitoneal shunt treatment.Copyright © 2016 Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.