- Gustav Andreisek, Lawrence M White, Yi Yang, Emma Robinson, Hai-Ling Margaret Cheng, and Marshall S Sussman.
- Department of Medical Imaging, Mount Sinai Hospital and the University Health Network, University of Toronto, Toronto, Ontario, Canada. gustav@andreisek.de
- Radiology. 2009 Sep 1; 252 (3): 865-73.
PurposeTo develop and verify the accuracy of a rapid imaging protocol for delayed gadolinium-enhanced magnetic resonance (MR) imaging of cartilage that was based on three-dimensional (3D) spoiled gradient-recalled acquisition in the steady state (SPGR) sequences with variable flip angles (FAs) (VFAs) and where a correction method for B(1) field inhomogeneities was applied.Materials And MethodsThe institutional research ethics board approved this study. Written informed consent was obtained from all subjects. A B(1) field inhomogeneity correction method was applied to a 3D SPGR pulse sequence with VFAs (repetition time msec/echo time msec, 7.1/3.3; FAs, 2 degrees , 5 degrees , 10 degrees , and 20 degrees ) and was used to perform delayed gadolinium-enhanced MR imaging of cartilage 3D T1 measurements at 1.5 T. The 3D T1 measurements were validated with the reference standard (the results of T1 mapping by using a single-section two-dimensional [2D] inversion-recovery [IR] fast spin-echo [SE] pulse sequence in vitro and in vivo) in six healthy volunteers.ResultsT1 values calculated from 3D T1 maps were not significantly different from reference T1 values in vitro (P = .195) and in vivo (P = .52) when a B(1) field inhomogeneity correction method was applied. In vivo T1 mapping of the articular surface of the whole femoropatellar joint, including data acquisition, was performed in approximately 8 minutes of acquisition time at a spatial resolution of 0.55 x 0.55 x 3.00 mm.ConclusionRapid T1 mapping by using 3D SPGR acquisitions with a VFA approach and a correction for B(1) field inhomogeneities can be used for delayed gadolinium-enhanced MR imaging of cartilage. T1 measurements performed in vitro and in vivo by using this approach are highly accurate when compared with those performed by using standard 2D IR fast SE T1 mapping as a reference.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.