    Reduction of inappropriate anti-tachycardia pacing therapies and shocks by a novel suite of detection algorithms in heart failure patients with cardiac resynchronization therapy defibrillators: a historical comparison of a prospective database.

    • Maurizio Lunati, Alessandro Proclemer, Giuseppe Boriani, Maurizio Landolina, Emanuela Locati, Roberto Rordorf, Elisabetta Daleffe, Renato Pietro Ricci, Domenico Catanzariti, Luca Tomasi, Michele Gulizia, Maria Stella Baccillieri, Giulio Molon, Maurizio Gasparini, and ClinicalService Cardiological Centres.
    • 'A De Gasperis' Cardiac Department, Niguarda Ca' Granda Hospital, Piazza Ospedale Maggiore, 3, 20162 Milano, Italy maurizio.lunati@ospedaleniguarda.it.
    • Europace. 2016 Sep 1; 18 (9): 1391-8.

    AimsImplantable cardioverter defibrillators improve survival of patients at risk for ventricular arrhythmias, but inappropriate shocks occur in up to 30% of patients and have been associated with worse quality of life and prognosis. In heart failure patients with cardiac resynchronization therapy defibrillators (CRT-Ds), we evaluated whether a new generation of detection and discrimination algorithms reduces inappropriate shocks.Methods And ResultsWe analysed 1983 Medtronic CRT-D patients (80% male, 67 ± 10 years), 1368 with standard devices (Control CRT-D) and 615 with new generation devices (New CRT-D). Expert electrophysiologists reviewed and classified the electrograms of all device-detected ventricular tachycardia/fibrillation episodes. Total follow-up was 3751 patients-years. Incidence of inappropriate shocks at 1 year was 2.8% [95% confidence interval (CI) = 2.0-3.5] in Control CRT-D and 0.9% (CI = 0.4-2.2) in New CRT-D (hazard ratio = 0.37, CI = 0.21-0.66, P < 0.001). In New CRT-D, inappropriate shocks were reduced by 77% [incidence rate ratio (IRR) = 0.23, CI = 0.16-0.35, P < 0.001] and inappropriate anti-tachycardia pacing by 81% (IRR = 0.19, CI = 0.11-0.335, P < 0.001). Annual rate per 100 patient-years for appropriate VF detections was 3.0 (CI = 2.1-4.2) in New CRT-D and 3.2 (CI = 2.1-5.0) in Control CRT-D (P = 0.68), for syncope was 0.4 (CI = 0.2-0.9) in New CRT-D and 0.7 (CI = 0.5-1.0) in Control CRT-D (P = 0.266), and for death was 1.0 (CI = 0.6-1.6) in New CRT-D and 3.5 (CI = 3.0-4.1) in Control CRT-D (P < 0.001).ConclusionDetection and discrimination algorithms used in new generation CRT-D significantly reduced inappropriate shocks when compared with standard CRT-D. This result, with no compromise on VF sensitivity or risk of syncope, has important implications for patients' quality of life and prognosis.Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: journals.permissions@oup.com.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.