    Effect of radiation therapy on permanent pacemaker and implantable cardioverter-defibrillator function.

    • Akash Makkar, Joann Prisciandaro, Sunil Agarwal, Morgan Lusk, Laura Horwood, Jean Moran, Colleen Fox, James A Hayman, Hamid Ghanbari, Brett Roberts, Diego Belardi, Rakesh Latchamsetty, Thomas Crawford, Eric Good, Krit Jongnarangsin, Frank Bogun, Aman Chugh, Hakan Oral, Fred Morady, and Frank Pelosi.
    • Department of Cardiac Electrophysiology, University of Michigan, Ann Arbor, Michigan, USA.
    • Heart Rhythm. 2012 Dec 1; 9 (12): 1964-8.

    BackgroundRadiation therapy's (RT's) effects on cardiac implantable electronic devices (CIEDs) such as implantable cardioverter-defibrillators (ICDs) and pacemakers (PMs) are not well established, leading to device removal or relocation in preparation for RT.ObjectiveTo determine the effect of scattered RT on CIED performance.MethodsWe analyzed 69 patients--50 (72%) with PMs and 19 (28%) with ICDs--receiving RT at the University of Michigan. Collected data included device model, anatomic location, and treatment beam energies, treatment type, and estimated dose to the device. Patients were treated with either high-energy (16-MV) and/or low-energy (6 MV) photon beams with or without electron beams (6-16 MeV). The devices were interrogated with pre- and post-RT and/or weekly with either in-treatment or home interrogation, depending on the patient's dependence on the device and the estimated or measured delivered dose. Outcomes analyzed were inappropriate ICD therapies, device malfunctions, or device-related clinical events.ResultsThe PMs were exposed to 84.4 ± 99.7 cGy of radiation, and the ICDs were exposed to 92.1 ± 72.6 cGy of radiation. Two patients with ICDs experienced a partial reset of the ICD with the loss of historic diagnostic data after receiving 123 and 4 cGy, respectively. No device malfunction or premature battery depletion was observed at 6-month follow-up from RT completion.ConclusionsCIED malfunction due to indirect RT exposure is uncommon. Regular in-treatment or home interrogation should be done to detect and treat these events and to ensure that diagnostic data are preserved.Copyright © 2012 Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.