• Pacing Clin Electrophysiol · Mar 2015

    Risk of device malfunction in cancer patients with implantable cardiac device undergoing radiotherapy: a population-based cohort study.

    • Tomas Zaremba, Annette Ross Jakobsen, Mette Søgaard, Anna Margrethe Thøgersen, Martin Berg Johansen, Laerke Bruun Madsen, and Sam Riahi.
    • Department of Cardiology, Center for Cardiovascular Research, Aalborg University Hospital, Aalborg, Denmark.
    • Pacing Clin Electrophysiol. 2015 Mar 1; 38 (3): 343-56.

    BackgroundPacemakers (PMs) and implantable cardioverter defibrillators (ICDs) may develop malfunction during external beam radiotherapy (RT). We aimed to describe clinical practice in PM/ICD patients undergoing RT and to assess the rate and predictors of device malfunctions.MethodsWe reviewed medical records of all PM/ICD patients undergoing RT at four centers in Western Denmark during 2003-2012. Logistic regression was applied to identify predictors of PM/ICD malfunctions.ResultsFive hundred sixty patients were included. The annual rate of RT courses in PM/ICD patients increased by 199% from 1.45 treatments per 100,000 person-years in 2003 to 4.33 in 2012. Safety measures included supplementary evaluations of PM/ICD (38.3%), reprogramming (1.5%), relocation of the device (3.5%), and application of a magnet to the ICD during RT (10.8%). At device evaluations after the RT (n = 453), malfunctions were detected in 10 (2.5%) PMs and four (6.8%) ICDs. Electrical resets constituted 11 (78.6%) of the malfunctions, and no failures were life-threatening or warranted PM/ICD removal. Factors associated with device malfunctions in logistic regression analysis were beam energy ≥15 MV (odds ratio [OR] 5.73; 95% confidence interval [CI], 1.58-20.76) and location of tumor below the diaphragm (OR 4.31; 95% CI, 1.42-13.12). However, the effect of tumor location declined (OR 2.27; 95% CI, 0.65-7.95) after adjustment for beam energy.ConclusionsAlthough the rate of RT in PM/ICD patients is increasing, the damaging effects of RT on the devices seem to be usually transient. Our data suggest that high beam energy plays the pivotal role in inducing impairments in these devices.©2015 Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.