• Europace · Apr 2014

    Comparative Study

    The effect of radiotherapy beam energy on modern cardiac devices: an in vitro study.

    • Tomas Zaremba, Annette Ross Jakobsen, Anna Margrethe Thøgersen, Lars Oddershede, and Sam Riahi.
    • Department of Cardiology, Center for Cardiovascular Research, Aalborg University Hospital, Hobrovej 18-22, 9100 Aalborg, Denmark.
    • Europace. 2014 Apr 1; 16 (4): 612-6.

    AimsRadiotherapy (RT) for malignancies can harm pacemakers (PMs) and implantable cardioverter-defibrillators (ICDs). There is some evidence that, besides cumulative dose, the damaging radiation effects increase with beam energy. The aim of this study was to determine whether modern PMs and ICDs are more sensitive to high-energy than to low-energy photon beams.Methods And ResultsTwo groups of unused PMs and explanted ICDs (five PMs and one ICD in each) were subjected to irradiations in a phantom with 6 and 18 megavolt (MV) photons, respectively. The devices were exposed to radiation at doses of 2 gray (Gy) daily to simulate two clinical scenarios with the PM/ICD in the RT field. A cumulative dose of 150 Gy was given to each device, corresponding to approximately twice the therapeutic dose. In the 6 MV group, one episode of PM malfunction was detected after reaching 150 Gy. In the 18 MV group, a total of 14 episodes of malfunction were detected starting at 30 Gy in all five PMs. No episodes appeared in the ICD, at the respective treatment groups. This corresponded to a hazard ratio of 9.11 [∼95% confidence interval (CI): 1.04-79.69] by Cox regression analysis between the two groups. In a repeated measures logistic regression model comparing the incidence rate of malfunctions, the odds ratio was 18.29 (∼95% CI: 1.52-219.41).ConclusionPhoton beam energy plays a considerable role in inducing implantable cardiac device malfunctions. Low-energy RT may be safer in PM/ICD patients despite relatively high radiation dose to the device.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.