- Pippa Storey, Fred J Frigo, R Scott Hinks, Bryan J Mock, Bruce D Collick, Nicole Baker, Jonathan Marmurek, and Simon J Graham.
- Radiology Department, Evanston Northwestern Healthcare, Evanston, Illinois 60201, USA. pstorey@enh.org
- Magn Reson Med. 2007 Mar 1; 57 (3): 614-9.
AbstractPartial k-space sampling is frequently used in single-shot diffusion-weighted echo-planar imaging (DW-EPI) to reduce the TE and thereby improve the SNR. However, it increases the sensitivity of the technique to bulk rotational motion, which introduces a phase gradient across the tissue that shifts the echo in k-space. If the echo is displaced into the high spatial frequencies, conventional homodyne reconstruction fails, causing intensity oscillations across the image. Zero-padding, on the other hand, compromises the image resolution and may cause truncation artifacts. We present an adaptive version of the homodyne algorithm that detects the location of the echo in k-space and adjusts the center and width of the homodyne filters accordingly. The adaptive algorithm produces artifact-free images when the echo is shifted into the high positive k-space range, and reduces to the standard homodyne algorithm in the absence of bulk motion.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.