Comparative Study
Single versus dual orthogonal plating for comminuted midshaft clavicle fractures: a biomechanics study.
- Glenn N Boyce, Andrew J Philpott, David C Ackland, and Eugene T Ek.
- Melbourne Orthopaedic Group, 33 The Avenue, Windsor, Melbourne, VIC, 3181, Australia.
- J Orthop Surg Res. 2020 Jul 9; 15 (1): 248.
BackgroundDual orthogonal plating of clavicle fractures may provide greater stiffness and strength than unilateral plate constructs and allow the use of lower-profile plates. We aim to biomechanically compare three clavicle plating constructs in a comminuted clavicle fracture model.MethodFifteen clavicle sawbones were osteotomised, simulating a comminuted midshaft fracture and allocated to either: group 1, single superior plate (3.5 mm superior plate); group 2, combination plating (3.5 mm superior plate, 2.8 mm anterior plate) and group 3, dual mini-plates (two 2.8-mm orthogonal mini-plates). Specimens were biomechanically tested under torsion and cantilever bending. Construct stiffness (Nm/degree) and load to failure (Nm) were measured.ResultsGroup 2 had higher torsional (0.70 vs. 0.60 Nm/deg, p = 0.017) and cantilever bending stiffness (0.61 vs. 0.51 Nm/deg, p = 0.025) than group 1. Group 3 had lower cantilever bending stiffness (0.39 vs. 0.51 Nm/deg, p < 0.004) and load to failure (40.87 vs. 54.84 Nm, p < 0.01) than group 1. All dual plate constructs that catastrophically failed did so from fracture at the lateral ends of the plates. Single plate constructs failed due to plate bending.ConclusionDual orthogonal fixation with mini-plates demonstrated lower stiffness and strength than traditional superior plating. The addition of an anterior mini-plate to a traditional superior plating improved construct stiffness and may have a role in patients seeking early return to activity.Level Of EvidenceBasic science biomechanical study.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.