- Benjamin Gauter-Fleckenstein, Julia Nguyen, Lennart Jahnke, Timo Gaiser, Boris Rudic, Sylvia Büttner, Frederik Wenz, Martin Borggrefe, and Erol Tülümen.
- Department of Radiation Oncology, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, Germany. Electronic address: benjamin.gauter@umm.de.
- Radiother Oncol. 2020 Nov 1; 152: 196-202.
Background And PurposeProviding evidence for radiotherapy (RT)-induced effects on cardiac implantable electric devices (CIEDs) with focus on flattening filter free-volumetric modulated arc therapy (FFF-VMAT) at 6 and 10 MV as well as 3D-conformal radiotherapy (3D-CRT) at 18 MV.Materials And Methods68 CIEDs (64 implantable cardioverter-defibrillators (ICDs) and 4 cardiac pacemakers (PMs)) were located on the left chest position on a slab phantom and irradiated under telemetrical surveillance either directly, or distant to 3D-CRT or FFF-VMAT, dose-rate 2500 cGy/min, and target dose of 150 Gy. Devices were placed within, close by (2.5 cm and 5 cm), and distant (35 cm) to the radiation field. Scatter radiation (SR) and photon neutrons (PN) were recorded. CIEDs were investigated in following groups: 1a) 18 MV 3D-CRT - 4 ICDs/4 PMs out of radiation field, 1b) 18 MV open field - 4 ICDs/4 PMs within radiation field, 2) 6 MV FFF-VMAT, 15 ICDs in 35 cm distance to VMAT, 3) 10 MV-FFF VMAT, 15 ICDs in 35 cm distance to VMAT, 4) 6 MV FFF-VMAT, 15 ICDs in 2.5 cm distance to VMAT, 5) 10 MV FFF-VMAT, 15 ICDs in 2.5 cm distance to VMAT.ResultsNo incidents occurred at 6 MV FFF. 10 MV FFF-VMAT and 18 MV 3D-CRT resulted in data loss, reset, and erroneous sensing with inhibition of pacing (leading to inadequate defibrillation) in 8/34 ICDs and 2/4 PMs which were not located within radiation. Direct radiation triggered instantaneous defibrillation in 3/4 ICDs.Conclusions6 MV FFF-VMAT is safe even at high dose-rates of 2500 cGy/min, regardless whether CIEDs are located close (2.5 cm) or distant (35 cm) to the radiation beam. CIEDs should never be placed within radiation and energy should always be limited to 6 MV. At 6 MV, VMAT at high dose-rates can be used to treat tumors, which are located close to CIEDs.Copyright © 2019 Elsevier B.V. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.