• J Magn Reson Imaging · Dec 2004

    Comparative Study

    Comparative study of FAST gradient echo MRI sequences: phantom study.

    • Rosalind L Perrin, Marko K Ivancevic, Sebastian Kozerke, and Jean-Paul Vallee.
    • Radiology Department, University Hospital of Geneva, Genève, Switzerland.
    • J Magn Reson Imaging. 2004 Dec 1; 20 (6): 1030-8.

    PurposeTo investigate a balanced steady state free precession sequence (b-SSFP) under a large range of conditions and to compare its performance with other types of gradient echo sequences for dynamic imaging.Materials And MethodsBalanced turbo field echo (b-TFE; Philips Medical Systems, Best, The Netherlands) was investigated in vitro at a range of T2/T1 along with T1-contrast enhanced turbo field echo (T1-TFE) and turbo field echo (TFE) so that a comparison could be made. Performance was quantified in terms of the initial slope of the signal-to-noise ratio (SNR) vs. 1/T1 curve (sensitivity) and the range of 1/T1 before signal saturation (contrast dynamic range [CDR]).ResultsThe b-TFE sequence was found to best perform, in terms of an optimal CDR, with a 90 degrees flip angle (FA), saturation preparation, and short inversion time. Using these parameters, the sensitivity was also higher than that of the TFE sequence and T1-TFE sequence under their respective optimal conditions. For detection of small changes in contrast agent concentration (0.0-0.1 mM), b-TFE was also found to be the sequence of choice, with optimized parameters as follows, 90 degrees FA, shortest TR/TE, and no magnetization preparation. The smallest matrices gave the highest signal sensitivity for all three sequences.ConclusionThe CDR of b-TFE was much narrower than that of T1-TFE but could be widened under optimized conditions. The sensitivity of the b-TFE technique was the highest of the three sequences under all conditions tested.2004 Wiley-Liss, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…