Arch. Pathol. Lab. Med. · Apr 1990
Calibration as a source of imprecision in cholesterol testing. Ramifications for patient risk classification.
- H D Epstein, H C Nipper, and G L Horowitz.
- Department of Pathology, Charles A. Dana Research Institute, Boston, Mass.
- Arch. Pathol. Lab. Med. 1990 Apr 1; 114 (4): 399-402.
AbstractRoutine calibration of a cholesterol assay system may compromise rather than improve precision. We compared an enzymatic assay on a centrifugal analyzer using a fixed factor with a factor recalculated from the response of standards assayed with each run. Over 36 batch runs, using three quality control materials, we found no statistically significant difference between the two methods in mean value, but in every case the fixed factor values were significantly more precise. With the risk classification system in effect at the time of the study, 32 (9.4%) of 342 patient serum specimens assayed for cholesterol were classified differently based solely on the method of data reduction. Thus, recalibration of our cholesterol assay system contributed to greater imprecision and to discrepancies in classification of patients' risk levels.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.